Drugs & medicalClass I

Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore…

Read the official notice →
Date
2026-06-10
Country
United States
Brand
Fresenius Kabi
Product
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
Classification
Class I
Units
Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Fresenius Kabi recalls

  1. 2026-07-29
  2. US
    Drugs & medicalClass II

    Fresenius Kabi USA, LLC recalls Glycopyrrolate Injection, USP, 1 mg per 5 mL (0.2 mg per mL), 25 x 5 mL fill in a 6.5 mL vial, Rx only, Multiple Dose Vials, Fresenius…

    Presence of Particulate Matter: Hair was found in product

    FDA · DrugFresenius Kabi USA, LLCNationwide within in the USAOfficial notice →
  3. 2026-06-17
  4. US
  5. 2026-06-10
  6. US
    Drugs & medicalClass II

    Fresenius Kabi USA, LLC recalls The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Model Number: LVP-0004.

    Potential for the Cassette Loading Lever to break.

    FDA · DeviceFresenius Kabi USA, LLCUS Nationwide distribution in the states of CA, CO, FL, GA, ID, IL, MA, MD,…Official notice →
  7. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.

    Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.

    FDA · DeviceFresenius Kabi USA, LLCUS Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD,…Official notice →
  8. 2026-04-15
  9. US
    Drugs & medicalClass II

    Fresenius Kabi USA, LLC recalls 0.9% Sodium Chloride Injection, USP, (2,250 mg per 250 mL) (9 mg per mL) 250 mL in a 250 mL freeflex bag, Rx only, Fresenius Kabi USA,…

    Lack of Assurance of Sterility

    FDA · DrugFresenius Kabi USA, LLCUS Nationwide , Alaska, and Puerto Rico.Official notice →
  10. 2026-01-21
  11. US
    Drugs & medicalClass II

    Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), LVP Software: LVP-SW-0005.

    Emphasizing instructions for LVP duration programming located in the IFU.

    FDA · DeviceFresenius Kabi USA, LLCUS Nationwide distribution in the states of CA, CO, FL, GA, ID, IL, MD, MI,…Official notice →
All 31 Fresenius Kabi recalls →

Other drugs & medical recalls in the United States

  1. 2026-06-10
  2. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with…

    Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system…

    FDA · DeviceBaxter Healthcare CorporationWorldwide - US Nationwide distribution including in the states of Alabama,…Official notice →
  3. US
    Drugs & medicalClass II

    Covidien LLC recalls COVIDIEN Shiley Tracheostomy Tube Cuffed with Inner Cannula REF 6FEN UDI-DI code: 60884522007028 The devices are intended for use in…

    Due to customer complaint regarding incorrect display box labeling.

    FDA · DeviceCovidien LLCWorldwide - U.S.: Nationwide distribution in the state of TX and the countries…Official notice →
  4. US
    Drugs & medicalClass I

    Becton Dickinson & Company recalls BD¿Spinal Tray with BD¿Whitacre Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL…

    This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing…

    FDA · DeviceBecton Dickinson & CompanyNationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS,…Official notice →
  5. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.

    Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.

    FDA · DeviceFresenius Kabi USA, LLCUS Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD,…Official notice →
  6. US
    Drugs & medicalClass I

    KAYSERBETTEN GMBH & CO. KG recalls Pediatric care bed; Product Designation: KayserBett IDA;

    If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third…

    FDA · DeviceKAYSERBETTEN GMBH & CO. KGDistribution US Nationwide and Canada.Official notice →
  7. US
    Drugs & medicalClass I

    Draeger, Inc. recalls Atlan A350XL. Model Number: 8621600. anesthesia workstation

    The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

    FDA · DeviceDraeger, Inc.US distribution in WI, AR, MN, WI, CO, MA, PA, AZ, FL, NY, SD, ND.…Official notice →
See all drugs & medical recalls →