Encore Medical, LP recalls EMPOWR 3D Knee Tibial Insert, 6R 10MM, VE, REF: 342-10-706
Knee insert package may contain an acetabular system, which may cause surgical delay to get a new product.
Read the official notice →- Date
- 2025-01-01
- Country
- United States
- Brand
- Encore Medical
- Product
- EMPOWR 3D Knee Tibial Insert, 6R 10MM, VE, REF: 342-10-706
- Classification
- Class II
- Units
- US Nationwide distribution in the states of IN, KS, CA, FL, NY, TX, AL, WA.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Encore Medical recalls
- 2025-07-23
US
Encore Medical, LP recalls Brand Name: EMPOWR 3D KNEE TIBIAL INSERT - E-PLUS Product Name: EMPOWR 3D KNEETM INS, 11L 12MM, VE Model/Catalog Number: 341-12-711…
Due to packaging discrepancy with the incorrect tibial inserts/implants being packaged. The use of this product may result in the incorrect product being selected. This could result in a revision surgery if used.
FDA · DeviceEncore Medical, LPU.S Distribution to states of: AZ, IA, ID, IL, IN, LA, MI, MO, MS, NE, NY, OK,…Official notice →- 2025-07-02
US
Encore Medical, LP recalls Brand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm Model/Catalog Number: 506-04-126 Software Version:…
Package labeling for Altivate Reverse Torx Peripheral screw contains a different size then actual included product.
FDA · DeviceEncore Medical, LPAt time of quarantine: RFK Medical Inc. Virginia (VA) CE102 506-04-122…Official notice →- 2025-01-01
US
Encore Medical, LP recalls EMPOWR 3D Knee Tibial Insert, 5R 16MM, VE, REF: 342-16-705; EMPOWR 3D Knee Tibial Insert, 3R 14MM, VE, REF: 342-14-703
5R 16MM knee tibial insert package may contain knee tibial insert 3R 14MM and vice versa, which may cause surgical delay to get a new product.
FDA · DeviceEncore Medical, LPUS Nationwide distribution in the states of VA, MN, TX, WI, IA, IL, IN, SC,…Official notice →- 2024-01-10
US
Encore Medical, LP recalls Reverse Shoulder System e+, REF 509-01-032, RSP Humeral Socket Insert, Sz. 32mm, Semi-Constrained, Sterile H2o2, CE 2797
Mislabeling
FDA · DeviceEncore Medical, LPUS Nationwide distribution in the states of VA, TN, MN, IN, RI, ME, TX, LA,…Official notice →- 2023-12-20
US
Encore Medical, LP recalls RSP Glenoid Baseplate, 30mm P2 Porous Coated REF 508-32-204. DJO Surgical shoulder device.
Coating hex did not engage with the Straight Mod Hexdriver or with the Hex Head Screw Driver
FDA · DeviceEncore Medical, LPU.S.: AZ, CA, FL, IN, KS, KY, ME, MN, MS, NJ, OH, Puerto Rico, RI, SC, TN, TX,…Official notice →- 2023-12-13
US
Encore Medical, LP recalls DJO surgical EMPOWR Porous Knee System Empowr 3D Femur-Porous Coated SIZE 8 Left 3D MATRIX COATED REF 243-01-108 DJO surgical EMPOWR…
Due to incorrect product/device within packaging.
Other drugs & medical recalls in the United States
- 2025-01-01
US
Siemens Healthcare Diagnostics, Inc. recalls Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;
Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Encore Medical, LP recalls EMPOWR 3D Knee Tibial Insert, 5R 16MM, VE, REF: 342-16-705; EMPOWR 3D Knee Tibial Insert, 3R 14MM, VE, REF: 342-14-703
5R 16MM knee tibial insert package may contain knee tibial insert 3R 14MM and vice versa, which may cause surgical delay to get a new product.
FDA · DeviceEncore Medical, LPUS Nationwide distribution in the states of VA, MN, TX, WI, IA, IL, IN, SC,…Official notice →US
Moximed, Inc. recalls MISHA Knee System Implant Small, Left, REF: 2-1001. MISHA Knee System Implant Small, Right, REF: 2-1002.
Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks,…
FDA · DeviceMoximed, Inc.US Nationwide distribution in the states of CA, OR, NY, OH, TX, CO, NY, VA.Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls ST REPROCESSED BRUNS CURETTE #00 (BC00), Medline Item No. 67600K
Weak seals may result in a breach of sterility, which may not be detectable to the user. This can lead to increased risk of contamination and infection.
US
Abbott Medical recalls The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates…
Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the…
FDA · DeviceAbbott MedicalWorldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →US
SEASPINE ORTHOPEDICS CORPORATION recalls Brand Name: Daytona Small Stature Product Name: CoCr LockingCap 4.5mm, Small Stature, T25, BT Model/Catalog Number: 79-0002…
Due to manufacturing defects observed on Locking Caps that affect the ability for components to mate.
FDA · DeviceSEASPINE ORTHOPEDICS CORPORATIONU.S. Nationwide distribution in the states of CA, CO, FL, GA, PA, TX, and VA.Official notice →