Drugs & medicalClass II

Elekta, Inc. recalls MOSAIQ Oncology Information System

A drug strength in MOSAIQ can be changed during the ordering process and approved, but the saved drug strength reverts to the original value.

Read the official notice →
Date
2021-10-06
Country
United States
Brand
Elekta
Product
MOSAIQ Oncology Information System
Classification
Class II
Units
Distribution was made to CA, CO, FL, ID, LA, MO, MS, NC, NH, NM, NV, NY, OH,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Elekta recalls

  1. 2023-06-07
  2. US
    Drugs & medicalClass II

    Elekta, Inc. recalls Elekta Medical Linear Accelerator (ELMA) included in the below platforms: a. Elekta Infinity; b. Versa HD; c. Elekta Synergy

    A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601

    FDA · DeviceElekta, Inc.Worldwide distribution. US distribution to AR, IN, LA, OK, TN, VA, WA.…Official notice →
  3. 2023-03-15
  4. US
    Drugs & medicalClass II

    Elekta Inc recalls Monaco RTP System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

    Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation.

    FDA · DeviceElekta IncNationwide distribution in the US. International distribution to Albania,…Official notice →
  5. 2022-12-21
  6. US
    Drugs & medicalClass II

    Elekta, Inc. recalls Oncentra Brachy radiation therapy planning system software.

    A reconstruction error may occur when using the Catheter Bending functionality in Applicator Modeling or Implant Modeling

    FDA · DeviceElekta, Inc.Distribution US nationwide, including Puerto Rico. There was no…Official notice →
  7. 2021-11-03
  8. US
    Drugs & medicalClass II

    Elekta, Inc. recalls Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00

    Under certain circumstances, there can be a difference between the planned and delivered dose of the two ovoid channels.

    FDA · DeviceElekta, Inc.Distribution throughout USA OUS distribution including:Algeria, Argentina,…Official notice →
  9. 2021-10-20
  10. US
    Drugs & medicalClass II

    Elekta, Inc. recalls Elekta Medical Linear Accelerator (EMLA), configured as follows: a. Elekta Precise Digital Accelerator; b. Elekta Harmony; c. Elekta…

    There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

    FDA · DeviceElekta, Inc.Distributed nationwide to FL, NC, OH, CA, IL, MD, OR, AL, NY, PA, MA, AK, GA,…Official notice →
  11. 2021-10-06
  12. US
    Drugs & medicalClass II

    Elekta Inc recalls Elekta Monaco - Product Usage: used to make treatment plans for patients with prescriptions for external beam radiation therapy.

    Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.

    FDA · DeviceElekta IncWorldwide distribution - US Nationwide distribution in the states of PA, MI,…Official notice →
All 18 Elekta recalls →

Other drugs & medical recalls in the United States

  1. 2021-10-06
  2. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls Flow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI, GTIN 0080682102339; (2) Aestiva 7900; (3)…

    The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cause leaks resulting in incorrect anesthesia machine tidal volumes potentially leading to over-delivery of…

    FDA · DeviceGE Healthcare, LLCDistribution was nationwide, including PR. Foreign distribution was made to…Official notice →
  3. US
    Drugs & medicalClass II

    Shanghai United Imaging Healthcare Co., Ltd. recalls uEXPLORER PET/CT System - Product Usage: intended to be operated by qualified healthcare professionals to assist in…

    The EXAM acquisition workflow could occasionally freeze after canceling the PET scan.

    FDA · DeviceShanghai United Imaging Healthcare Co., Ltd.U.S. Nationwide distribution in the states of CA, IL, IN, TX, and WI.Official notice →
  4. US
    Drugs & medicalClass II

    Atrium Medical Corporation recalls Atrium Pneumostat Chest Drain Valve, Part Number 16100

    The firm is revising the Indication for Use, Contraindications, Warnings, and Precautions, after the device was involved in the treatment of a preterm infant who later died.

    FDA · DeviceAtrium Medical CorporationDomestic distribution nationwide. Foreign distribution worldwide.Official notice →
  5. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls AMX Navigate Mobile X-Ray System

    An unexpected column motion while attempting to park the column on the AMX Navigate mobile X-ray system.

    FDA · DeviceGE Healthcare, LLCDomestic Distribution OnlyOfficial notice →
  6. US
    Drugs & medicalClass II

    Atrium Medical Corporation recalls Atrium ADVANTA VXT Vascular Graft, Part No. 22059, UDI 00650862220599 - Product Usage: intended for use in arterial vascular…

    The product package may potentially contain a mismatched graft that does not contain the radial support ring (helix) identified in the labeling.

    FDA · DeviceAtrium Medical CorporationWorldwide distribution - US Nationwide distribution in the state of Florida…Official notice →
  7. US
    Drugs & medicalClass II

    ARJOHUNTLEIGH POLSKA Sp. z.o.o. recalls Arjo Citadel beds originally assembled with the IndiGo modules during manufcturing, model number CX812A3J3AMAD0; and IndiGo…

    The IndiGo power cord might wear during use which could lead to its damage, resulting in the cord being cut, burnt or short circuit with limited emission of sparks.

    FDA · DeviceARJOHUNTLEIGH POLSKA Sp. z.o.o.US Nationwide distribution in the states of CA, FL, IL, KY, MD, NV, NY, PA,…Official notice →
See all drugs & medical recalls →