Drugs & medicalClass II

Dynex Technologies, Inc. recalls Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility…

The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible…

Read the official notice →
Date
2025-08-27
Country
United States
Brand
Dynex Technologies
Product
Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility Automated…
Classification
Class II
Units
Domestic: CA, FL, NC, NJ, NY;
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Dynex Technologies recalls

  1. 2022-06-08
  2. US
    Drugs & medicalClass II

    Dynex Technologies, Inc. recalls DYNEX Agility, Agility Analyzer, Model No. 67000

    Control samples aspirated from wrong SmartKit on the Agility. Agility software was updated to v1.4.7 to resolve the issue. This leads to a risk that a control from another assay's SmartKit will be used instead of the…

    FDA · DeviceDynex Technologies, Inc.Worldwide distribution - US Nationwide distribution in the states of AL, CA,…Official notice →
All 2 Dynex Technologies recalls →

Other drugs & medical recalls in the United States

  1. 2025-09-03
  2. US
    Drugs & medicalClass II

    Amerisource Health Services LLC recalls Sucralfate Tablets, USP 1 gram, 100 Tablets, (10x10), Rx Only, Distributed by: American Health Packaging, Columbus, Ohio 43217…

    CGMP Deviations: The recalling firm filed for Chapter 11 in September 2024. As a result, it cannot monitor the quality program and hence cannot assure that products meet the identity, strength, quality, and purity…

    FDA · DrugAmerisource Health Services LLCNationwideOfficial notice →
  3. US
    Drugs & medicalClass II

    Cardinal Health Inc. recalls Enoxaparin Sodium Injection, USP, 30 mg/0.3 mL, packaged in 0.3 mL prefilled syringes further packaged in bags of 5 prefilled syringes, Rx…

    Correct Labeled Product Mispack: Bags labeled for Enoxaparin Sodium Injection, 80 mg/0.8 mL, contained Enoxaparin Sodium Injection, 30 mg/0.3 mL

    FDA · DrugCardinal Health Inc.KSOfficial notice →
  4. US
  5. 2025-08-27
  6. US
    Drugs & medicalClass II

    LUMENIS, LTD. recalls OtoLase Starter Kit; UPN: AC-2003748. Indicated for a variety of surgical uses including ablation, coagulation, incision, excision, and…

    The potential for unsterilized product within finished product labeled as sterile.

    FDA · DeviceLUMENIS, LTD.Domestic: PA, WI; International: Germany, India;Official notice →
  7. US
    Drugs & medicalClass II

    Boston Scientific Neuromodulation Corporation recalls Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep…

    Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural…

    FDA · DeviceBoston Scientific Neuromodulation CorporationWorldwide - US Nationwide distribution including in the states of MN, FL, CO,…Official notice →
  8. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP).…

    Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →