DRG International, Inc. recalls Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description: T4 Total ELISA
An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2 months (i.e., 2026-03-31…
Read the official notice →- Date
- 2025-08-13
- Country
- United States
- Brand
- DRG
- Product
- Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description: T4 Total ELISA
- Classification
- Class III
- Units
- Worldwide - US Nationwide distribution in the states of CA, MA, PA and the…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More DRG recalls
- 2026-10-14
US
DRG International, Inc. recalls Sperm Ab ELISA; Model/Catalog Number: EIA-1826; Product Description: Sperm Ab ELISA (96 well plate test);
Potential for false negative patient results.
FDA · DeviceDRG International, Inc.Worldwide - US Nationwide distribution in the state of NJ and the countries of…Official notice →- 2025-01-22
US
DRG International, Inc. recalls DRG:HYBRiD-XL Insulin. Model/Catalog Number: HYE-5361
The user of the assay may either face a failure message during re-calibration of the Hybrid.XL instrument using the re-calibration R1 and R2 of the device or find controls out of specification (failed Control High)…
FDA · DeviceDRG International, Inc.Worldwide - US Nationwide distribution in the state of New Jersey and the…Official notice →- 2024-11-06
US
DRG International, Inc. recalls 17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone…
The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to…
FDA · DeviceDRG International, Inc.AZ, CA, NH Foreign: Austria Belgium Georgia Germany Greece India Irak Italy…Official notice →- 2024-08-21
US
DRG International, Inc. recalls Product Name: Salivary Testosterone ELISA Model/Catalog Number: SLV-3013 Product Description: An enzyme immunoassay for the quantitative…
Invalid Controls. The Low Control is out of range; too high which causes invalid runs that would result in repeating the runs and delays for reporting results to patients.
FDA · DeviceDRG International, Inc.US Nationwide distribution in the states of CA, IL, MN, PA, UT, VA.Official notice →- 2022-04-13
US
DRG International, Inc. recalls C-Peptide ELISA : IVD measurement of C-Peptide in serum, plasma (EDTA-, heparin- or citrate plasma) and urine. Catalog # EIA-1293
ODs of the standards too low(< 0.6 ODmax) may cause invalid runs, delayed patient results are possible
- 2021-12-08
US
DRG International, Inc. recalls Total Thyroxine (T4) ELISA; REF EIA-1781
Firm has identified two lots of product that do not meet product shelf life, resulting in failed Controls. The clinical impact may be a delay in diagnosis and a consumption of patient samples.
FDA · DeviceDRG International, Inc.Domestic: NY; International: TZ, VN, BD, CO, TT, YE, BO, DE, SV, GE, PY, TG,…Official notice →
Other drugs & medical recalls in the United States
- 2025-08-13
US
Nuclein LLC recalls DASH SARS-CoV-2 & Flu A/B Test Model/Catalog Number: SG-0006 combination COVID-19 and Flu test. Each DASH SARS-CoV-2 & Flu A/B Test is…
Due to manufacturing error, assay test may result in missed diagnosis or delay of treatment. (Potential false negative result)
FDA · DeviceNuclein LLCUS distribution to: Florida, New York, Texas, and West VirginiaOfficial notice →US
Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L.…
FDA · DeviceSiemens Healthcare Diagnostics IncWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →US
Beckman Coulter, Inc. recalls DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current…
US
Beckman Coulter Inc. recalls UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,
Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL…
FDA · DeviceBeckman Coulter Inc.US: AK, AL, AR, CO, CT, DC, DE, ID, UT OUS: Algeria, Argentina, Australia,…Official notice →US
AVID Medical, Inc. recalls GYN ROBOTIC PACK. Medical convenience kit.
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
US
Philips Medical Systems (Cleveland) Inc recalls Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software
Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,
FDA · DevicePhilips Medical Systems (Cleveland) IncWorldwide - US Nationwide distribution in the state of WV and the countries of…Official notice →