Drugs & medicalClass II

DeRoyal Industries Inc recalls DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement

The tubing may not ensure that the yankauer stays intact, and secure to the tubing / yankauer connection for a tight seal. This could result in inadequate suction and cause delay in patient care.

Read the official notice →
Date
2025-02-26
Country
United States
Brand
DeRoyal Industries
Product
DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement
Classification
Class II
Units
U.S.: ND, NY, VA, WV, NJ, TN, OK, TN, GA, OH, MN, AZ, WI, IL, MA, MD, MO, AR,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More DeRoyal Industries recalls

  1. 2026-09-30
  2. US
    Drugs & medicalClass II

    DeRoyal Industries Inc recalls REF: 81-030415 Myocardial Probe Non-Pyrogenic , 400 Series

    Myocardial temperature probes contain an incorrect needle. Specifically, an 18mm needle was included instead of a 15mm needle.

    FDA · DeviceDeRoyal Industries IncUS Nationwide distribution in the states of LA,MO, MD, TN, UT.Official notice →
  3. 2026-06-10
  4. US
    Drugs & medicalClass II

    DeRoyal Industries Inc recalls DeRoyal Industries Inc. Ref 46-406, Covaderm Plus, Vascular Access Dressing 6"x6" (15cmx15cm)

    Sterile wound dressing, lacks sterility assurance

    FDA · DeviceDeRoyal Industries IncUS Nationwide distribution in the states of FL & NC.Official notice →
  5. 2025-03-12
  6. US
    Drugs & medicalClass II

    DeRoyal Industries Inc recalls DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing,…

    Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a…

    FDA · DeviceDeRoyal Industries IncU.S.: IL, MN, NC, PA, and VA O.U.S.: N/AOfficial notice →
  7. 2024-11-13
  8. US
    Drugs & medicalClass II

    DeRoyal Industries Inc recalls DeRoyal CHILDREN'S LAPTRACEPACK, 89-9362.06

    DeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual…

    FDA · DeviceDeRoyal Industries IncUS Nationwide distribution in the state of OH.Official notice →
  9. 2024-04-03
  10. US
    Drugs & medicalClass I

    DeRoyal Industries Inc recalls GeoMed Custom Tracecarts, REF Numbers: a) 53-1831 ORTHO TOTAL JOINT TRACECART, b) 53-1836 GENERAL SURGERY TRACECART

    DeRoyal manufactured and distributed Surgical Tracecarts that contain a Nurse Assist 16FR Silver Urine Meter Foley which were subsequently recalled.

    FDA · DeviceDeRoyal Industries IncFloridaOfficial notice →
  11. 2024-01-03
  12. US
    Drugs & medicalClass II

    DeRoyal Industries Inc recalls DeRoyal CARDIAC CATH SET-UP KIT PGYBK, REF 89-10572

    Kit were manufactured and distributed with Edwards Life Sciences TruWave Disposable pressure transducers which were subsequently recalled by Edwards Life Sciences.

    FDA · DeviceDeRoyal Industries IncUS: VirginiaOfficial notice →
All 18 DeRoyal Industries recalls →

Other drugs & medical recalls in the United States

  1. 2025-02-26
  2. US
    Drugs & medicalClass II

    Hollister Incorporated recalls AnchorFast SlimFit, Oral Endotracheal Tube Fastener, REF 9787,

    Hollister received reports of decreased skin barrier wear time which could can lead to tube migration

    FDA · DeviceHollister IncorporatedWorldwide distribution - US Nationwide and the country of Canada.Official notice →
  3. US
    Drugs & medicalClass II

    XTANT Medical Holdings, Inc recalls Irix-A Integrated Lumbar Fusion System, Model Number X080-423011-08PC (Spacer Assy, 42 X 30 X 8¿, 11mm, PC (STERILE)) and Model…

    Certain implants in the Irix-A Lumbar Fusion System were distributed while labeled with an incorrect expiration date.

    FDA · DeviceXTANT Medical Holdings, IncWorldwide - US Nationwide distribution in the states of AZ, CA, CO, LA, OH,…Official notice →
  4. US
    Drugs & medicalClass II

    KCI USA, INC. recalls 3M V.A.C. VIA Negative Pressure Wound Therapy System REF: VIAKIT077D01/US - V.A.C. VIA 7 Day Kit VIAKIT07S05/AU - V.A.C. VIA Starter Kit…

    Due to increase in complaints related to leak alarms

    FDA · DeviceKCI USA, INC.Worldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →
  5. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Medline RR-LIVER TRNS ACCESSORY PEDS, DUNJ63397; medical procedure convenience kit

    Medline is recalling medical convenience kits which were assembled with specific Coloplast Foley catheters and Prostatic catheters which were subsequently recalled due to a potential sterility issue detected during…

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS: CAOfficial notice →
  6. US
    Drugs & medicalClass II

    Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming…

    Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .

    FDA · DeviceAugustine Temperature Management, LLCUS distribution to states of: AL, TX, NVOfficial notice →
  7. US
    Drugs & medicalClass I

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane; Catalog numbers: (1) 722008, (2) 722013.

    Possibility of the patient falling from the table related to the mattress used on the patient table.

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldiwde distribution - US Nationwide and the countries of Afghanistan,…Official notice →