Drugs & medicalClass II

DePuy Orthopaedics, Inc. recalls VELYS Robotic-Assisted Solution Base Product No.: 451570100

System software v1.5.1 has a system software issue related to the Daylight Savings Time (DST) change that can cause a system error, requiring the user to restart the system and potentially cause a delay in treatment

Read the official notice →
Date
2022-05-11
Country
United States
Brand
DePuy Orthopaedics
Product
VELYS Robotic-Assisted Solution Base Product No.: 451570100
Classification
Class II
Units
Worldwide distribution - US Nationwide and the countries of Israel, Australia…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More DePuy Orthopaedics recalls

  1. 2023-03-29
  2. US
    Drugs & medicalClass II

    DePuy Orthopaedics, Inc. recalls Attune Revision Limb Preservation System (LPS) Tibial Insert XXSM 18MM, Part Number 151760118

    The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to…

    FDA · DeviceDePuy Orthopaedics, Inc.Worldwide distribution - US Nationwide and the countries of Canada, Poland,…Official notice →
  3. 2022-07-27
  4. US
    Drugs & medicalClass II

    DePuy Orthopaedics, Inc. recalls ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant…

    Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical…

    FDA · DeviceDePuy Orthopaedics, Inc.Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil,…Official notice →
  5. 2022-07-20
  6. US
    Drugs & medicalClass II

    DePuy Orthopaedics, Inc. recalls MODULAR CATHCART FRACTURE HEAD HIP BALL, sizes 41mm to 60mm inclusive; metallic femoral head prosthesis, Product No. 136341000,…

    The Cathcart Ball has an inherent +5mm offset which is not currently specified in its Instructions for Use or Surgical Technique Guide. The STG has been updated to reflect the clarification that the +5mm offset should…

    FDA · DeviceDePuy Orthopaedics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  7. 2022-02-16
  8. US
    Drugs & medicalClass II

    DePuy Orthopaedics, Inc. recalls ATTUNE REVISION POSTERIOR FEMORAL AUGMENT SZ 9 12MM-intended as a Component in total knee arthroplasty (TKA) revision surgeries Product…

    May have adhesive residue on the posterior surface of the femoral augment, if undetected prior to implantation, soft tissue irritation and/or infection may occur.

    FDA · DeviceDePuy Orthopaedics, Inc.US Nationwide distribution in the states of TN, TX.Official notice →
  9. 2022-01-19
  10. US
    Drugs & medicalClass II

    DePuy Orthopaedics, Inc. recalls GLOBAL UNITE REV STEM SZ 10 Product Code: 110010600

    Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage

    FDA · DeviceDePuy Orthopaedics, Inc.Worldwide distribution - US Nationwide and the countries of Austria,…Official notice →
  11. 2021-10-13
  12. US
    Drugs & medicalClass II

    DePuy Orthopaedics, Inc. recalls ATTUNE Revision Tibial Augment Size 7/8 5mm- intended for knee joint replacement Part Code:152307001

    Packaging may include screws that are not intended to be used with the 5mm augments, may cause delay in surgery

    FDA · DeviceDePuy Orthopaedics, Inc.US Nationwide distribution in the states of CA, LA, MN, MO, NC, NE, OR, PA,…Official notice →
All 8 DePuy Orthopaedics recalls →

Other drugs & medical recalls in the United States

  1. 2022-05-11
  2. US
    Drugs & medicalClass II

    Johnson & Johnson Surgical Vision Inc recalls TRAY RING COVER, REF OM271000, CONTAINS (1) TRAY RING COVER, STERILE R, Rx Only,

    Potential breach in the sterility barrier for tray ring covers.

    FDA · DeviceJohnson & Johnson Surgical Vision IncWorlwide distribution - US Nationwide distribution in the states of AL, AR,…Official notice →
  3. US
    Drugs & medicalClass II

    ASO LLC recalls First Aid Kits containing acetaminophen under brand names Equate (Walmart), Meijer, and Care Science (Amazon)

    Kits contained recalled acetaminophen.

    FDA · DeviceASO LLCUS Nationwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Acera Surgical Inc recalls Restrata, Part Number RWM1-2X2

    Misprinted expiration date on the shelf-box

    FDA · DeviceAcera Surgical IncNationwideOfficial notice →
  5. US
    Drugs & medicalClass II

    restor3d Inc. recalls Evans Wedge Disposable Trials, Medium Footprint, 10mm Thickness (6101-20181018). Orthopedic manual surgical instrument

    During re-work of disposable instruments to relabel for 4-year shelf-life of disposable instruments, parts got mixed into other lots.

    FDA · Devicerestor3d Inc.US Nationwide distribution in the states of KY, NC, NJ, TX, IN, VA and AL.Official notice →
  6. US
    Drugs & medicalClass II

    restor3d Inc. recalls Non-Sterile Double Ended Trial Kit, Small Footprint, (6110-N01-KTS01) for the Cervical Cage. The kit includes 3 trials. Orthopedic manual…

    Product labeled incorrectly. The lot was labeled with Small footprint when the product was Medium.

    FDA · Devicerestor3d Inc.US Nationwide distribution in the states of KY, NC, NJ, TX, IN, VA and AL.Official notice →
  7. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur Enhanced Estradiol (eE2) 500T - In vitro diagnostic use in the quantitative determination of estradiol SMN:…

    Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →
See all drugs & medical recalls →