Drugs & medicalClass II

Deerfield Imaging, Inc. recalls Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of the IMRIS HFD100 head fixation device, models…

Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may deflect (rotationally around the axis) and not return to within acceptable…

Read the official notice →
Date
2022-04-20
Country
United States
Brand
Deerfield Imaging
Product
Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of the IMRIS HFD100 head fixation device, models…
Classification
Class II
Units
Worldwide distribution - US Nationwide distribution in the states of AL, FL,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Deerfield Imaging recalls

  1. 2021-10-06
  2. US
    Drugs & medicalClass II

    Deerfield Imaging, Inc. recalls Operating room table - Product Usage: intended for use during diagnostic examinations, radiologic procedures or surgical procedures to…

    ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware issue.

    FDA · DeviceDeerfield Imaging, Inc.Worldwide distribution - US Nationwide distribution FL, NH, TX, OH, CA, OR,…Official notice →
All 2 Deerfield Imaging recalls →

Other drugs & medical recalls in the United States

  1. 2022-04-20
  2. US
    Drugs & medicalClass II

    Varian Medical Systems Imaging Laboratory GmbH recalls ARIA Radiation Therapy Management (RTM) VERSIONS 13.6, 15.1, 15.5, 15.6, 16.0, 16.1

    Software issue for treatment plan and image management application may result in mismatch values which could result in treatment in the wrong location

    FDA · DeviceVarian Medical Systems Imaging Laboratory GmbHWorldwide distribution - US Nationwide distribution in the states of MI, KS,…Official notice →
  3. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Medline Standard PVC Laryngeal Masks, 1 device per pouch, 5 each pouches per box

    The mask cuff may disconnect from the device's breathing tube.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Graphic Controls Acquisition Corporation recalls CLARAVUE Pre-wired Electrode - IP SET M3 RT ADULT-US Model: 32028777 50600 US

    Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without…

    FDA · DeviceGraphic Controls Acquisition CorporationUS Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN,…Official notice →
  5. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Stanbio TDM/Beta-Hydroxybutyrate Linearity Standards, packaged in boxes containing Levels 1-6, 1 x 4 mL each, REF 2450-604,…

    The product was stored incorrectly due to improper storage controls which may lead to delayed results.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldDistribution was made to AL, FL, NE OH, TX, WA, and WI. There was no…Official notice →
  6. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Hydroxybutyrate LiquiColor Tests - includes 3 bottles (1 of each size, 3mL, 8.5mL, and 50mL)

    Product was improperly store which may cause a delay in results.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS distribution only - AL, CA, FL, GA, IL, IN, LA, MO, MS, NC, NE, NY, OH, PA,…Official notice →
  7. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Fructosamine (Fruc), Siemens Material Number 11097637

    Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Global Distribution. US Nationwide.Official notice →