Covidien recalls Covideien McGRATH MAC Video Laryngoscope, REF: 301-000-000, Non-Sterile, RxOnly
IFU update to address device's battery handling information.
Read the official notice →- Date
- 2024-08-28
- Country
- United States
- Brand
- Covidien
- Product
- Covideien McGRATH MAC Video Laryngoscope, REF: 301-000-000, Non-Sterile, RxOnly
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Covidien recalls
- 2025-06-18
US
Covidien recalls Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT70M-WW-EU, HT70M-WW-NA, HT70-SY-EU.…
Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation,…
FDA · DeviceCovidienWorldwide - US Nationwide distribution including in the states of MA, CT, KS,…Official notice →- 2025-04-02
US
Covidien recalls Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R
Tracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration,…
FDA · DeviceCovidienOUS distribution only to countries of: France, Germany, Ireland, Italy,…Official notice →- 2024-11-06
US
Covidien LP recalls Chameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US), 6mm x 40mm x 75cm (CFN CH06-40-75US), 7mm x 40mm…
Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.
FDA · DeviceCovidien LPWorldwide Distribution: US (Nationwide) and OUS (foreign) to countries of:…Official notice →- 2024-08-28
US
Covidien recalls McGrath Mac Video Laryngoscope, REF: 300-000-000 , Non-Sterile, Rx Only
Battery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.
FDA · DeviceCovidienWorldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →- 2024-08-14
US
Covidien recalls Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600, PURITAN BENNETT PB560 VENTILATOR…
Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than…
FDA · DeviceCovidienUS: KY, OH, GA, MN, IN,UT, IL, NJ, NY, PR, LA, MD, MO, TN, MA, ME, CA, SC, AR,…Official notice →- 2024-06-19
US
Covidien, LLC recalls EsoFLIP, 30mm BALLOON DILATION CATHETER, REF ES-330, Rx Only
Due to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements that may lead to various harms.
FDA · DeviceCovidien, LLCWorldwide - US Nationwide distribution including in the states of AK, AL, AZ,…Official notice →
Other drugs & medical recalls in the United States
- 2024-09-04
US
Preferred Pharmaceuticals, Inc. recalls Ibuprofen Tablets, USP 400mg, Generic for Motrin, Pkg Size: 30 tablets per bottle, Mfg: Dr. Reddy's Laboratories, Louisiana,…
Failed Impurities/Degradation Specifications - at 18-month Stability testing
US
Dr. Reddy's Laboratories, Inc. recalls IBU Ibuprofen Tablets, USP, 800 mg, Rx Only, Distributor: Dr. Reddy's Laboratories Inc., Princeton, NJ 08540, Made in India, a)NDC…
Failed impurities/degradation specifications: results for unknown impurity, were 0.13% and 0.11% respectively, exceeding the 0.10% specification limit.
US
Glenmark Pharmaceuticals Inc., USA recalls Indomethacin Extended-Release Capsules, USP, 75 mg, packaged in a) 60-count bottle (NDC 68462-325-60) and b) 90-count bottle…
Failed Dissolution Specifications: below specification results
- 2024-08-28
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper systems (R8.1 with Poly-G Stand) Interventional Fluoroscopic X-Ray System, Model Numbers 722010,…
A half-threaded bolt was included in the LTE kits instead of the correct full-threaded bolt.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide distribution in the states of IL and OR. The…Official notice →US
NeuroLogica Corporation recalls GM60A. Digital Diagnostic Mobile X-Ray System.
Potential for the anti-fall system of the device arm to fail.
FDA · DeviceNeuroLogica CorporationUS Nationwide distribution in the states of AR, CT, GA, IN, KS, MA, MD, MO,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls LINER, SOFT, SUCTION CANISTER, 1500CC, REF DYNDSCL1500
A slight dimensional variation which has the potential for increased difficulty fitting into outer canister.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide - US Nationwide and the countries of Panama, Mexico, Canada, Qatar,…Official notice →