Drugs & medicalClass II

Covidien, LP recalls Tri-Staple 2.0 Black Intelligent Reload, Product Number SIG60AXT

Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These…

Read the official notice →
Date
2023-03-15
Country
United States
Brand
Covidien
Product
Tri-Staple 2.0 Black Intelligent Reload, Product Number SIG60AXT
Classification
Class II
Units
Worldwide distribution - US Nationwide distribution in the states of Arkansas,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Covidien recalls

  1. 2023-04-26
  2. US
    Drugs & medicalClass I

    Covidien recalls Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or Reusable Inner Cannula (REF,…

    A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories.…

    FDA · DeviceCovidienWorldwide - US Nationwide distribution including in the states of PR, CA, FL,…Official notice →
  3. 2023-04-19
  4. US
    Drugs & medicalClass II

    Covidien, LP recalls Step Bladeless Trocars, 10 mm cannula and dilator with radially expandable sleeve and 5 - 10 mm reducer, Product Number S101010

    Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of…

    FDA · DeviceCovidien, LPWorldwide Distribution: US (nationwide) to States of: AR, CA, FL, GA, LA, MA,…Official notice →
  5. 2023-04-05
  6. US
    Drugs & medicalClass II

    Covidien recalls COVIDIEN Shiley Lo-Pro Oral/Nasal Tracheal Tube Cuffed REF 86051 7.0 mm I.D. The Shiley Lo-Pro Oral/Nasal Tracheal Tube Cuffed are…

    Due to potential leaks associate with the Oral/Nasal Tracheal Tube Cuffed inflation system.

    FDA · DeviceCovidienU.S.: AK, AZ, CA, FL, GA, IA, IL, IN, KS MD, MI, MS, NC, NE, NJ, NY, OH, OK,…Official notice →
  7. 2023-03-08
  8. US
    Drugs & medicalClass II

    Covidien, LP recalls Parietex Composite Mesh Polyester with Absorbable Collagen Film, Horseshoe- Shaped, 9 x 8 cm (3.6" x 3.1"). Intended for the reinforcement…

    Collagen film was placed on the opposite side of the mesh may cause delay to treatment/therapy, adhesions, erosion/migration, pain, fistula, hernia (recurrence), and failure of implant

    FDA · DeviceCovidien, LPUS Distribution to states of: CA, GA, NY and OUS Foreign distribution to…Official notice →
  9. US
    Drugs & medicalClass II

    Covidien, LP recalls Surgilon Braided Nylon suture Product Description: 88861883-51 SURGILON* 2-0 BLK 75CM C16 88861915-71 SURGILON* 1 BLK 75CM GS11…

    Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

    FDA · DeviceCovidien, LPUS Nationwide Distribution and OUS Foreign distribution to countries of:…Official notice →
  10. 2023-01-18
  11. US
    Drugs & medicalClass II

    Covidien, LLC recalls COVIDIEN Emprint Ablation Catheter with Thermosphere Technology, Bronchoscopic Ablation, REF CA108L1

    The product is labeled with the incorrect expiration date.

    FDA · DeviceCovidien, LLCForeign Distribution: Hong KongOfficial notice →
All 53 Covidien recalls →

Other drugs & medical recalls in the United States

  1. 2023-03-15
  2. US
    Drugs & medicalClass II

    Elekta Inc recalls Monaco RTP System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

    Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation.

    FDA · DeviceElekta IncNationwide distribution in the US. International distribution to Albania,…Official notice →
  3. US
    Drugs & medicalClass II

    Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0800-35,…

    Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.

    FDA · DeviceDatascope Corp.Nationwide US distribution. Worldwide international distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-75, 0998-00-0800-83, and 0998-00-0800-85

    There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of…

    FDA · DeviceDatascope Corp.US Nationwide. Global Distribution.Official notice →
  5. US
    Drugs & medicalClass II

    Aniara Diagnostica LLC recalls BIOPHEN Protein C 5 kit is for in vitro quantitative determination of Protein C activity on human citrated plasma. Protein C is a…

    New warnings and precautionary statements on product labelling: Reagent 2 (R2) of BIOPHEN Protein C 5 contains the chemical substance - Cesium Chloride which changed classification and has become hazardous according to…

    FDA · DeviceAniara Diagnostica LLCUS Nationwide distribution in the state of NC.Official notice →
  6. US
    Drugs & medicalClass III

    Preat Corp recalls PREAT NOBIL GRIP, product numbers 0102000 and 01202000-3 (3 pack), 5-ml vials packaged in pouches, dental anaerobic glue.

    Product is labeled with an incorrect expiration date.

    FDA · DevicePreat CorpWorldwide - US Nationwide distribution including in the states of AL, AZ, CA,…Official notice →
  7. US
    Drugs & medicalClass II

    St. Jude Medical recalls CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA)…

    Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of…

    FDA · DeviceSt. Jude MedicalWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
See all drugs & medical recalls →