Cook Medical Incorporated recalls EchoTip¿ Ultra HD Endoscopic Ultrasound Access Needle, ECHO-HD-19-A, G52012
Devices may contain elevated levels of bacterial endotoxin.
Read the official notice →- Date
- 2024-05-22
- Country
- United States
- Brand
- Cook Medical
- Product
- EchoTip¿ Ultra HD Endoscopic Ultrasound Access Needle, ECHO-HD-19-A, G52012
- Classification
- Class II
- Units
- International distribution in the countries of Germany, Netherlands,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Cook Medical recalls
- 2025-10-22
US
Cook Medical Incorporated recalls Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W…
Affected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular embolization.
FDA · DeviceCook Medical IncorporatedDomestic distribution to AL AR AZ CA CT DC DE FL GA IL IN LA MA MD MI MN MO NC…Official notice →- 2024-01-24
US
Cook Medical Llc recalls MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429
The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional…
FDA · DeviceCook Medical LlcDomestic distribution to CA IN LA MD NY International distribution to…Official notice →- 2022-04-13
US
Cook Medical Incorporated recalls Gunther Tulip Vena Cava Filter Set for Femoral Vein Approach Catalog Number/GPN: IGTCFS-65-1-FEM-TULIP G52917; IGTCFS-65-1-JUG-TULIP…
IFU update includes an update to the Device description, Intended Use/Indications for Use, Contraindications, Warnings, Precautions, Potential adverse events
FDA · DeviceCook Medical IncorporatedUS Nationwide - Worldwide Distribution Foreign: Canada, Bahamas ,Chile GuamOfficial notice →
Other drugs & medical recalls in the United States
- 2024-05-29
US
SUN PHARMACEUTICAL INDUSTRIES INC recalls Xelpros (latanoprost ophthalmic emulsion) 0.005%, 125mcg/2.5mL, 2.5 mL bottle, Rx only, Manufactured by: Sun Pharmaceutical…
Failed Release Testing: Out of specification for particulate matter test.
- 2024-05-22
US
Cardinal Health 200, LLC recalls CardinalHealth Curity Flexible Adhesive Bandage X-Large 2 x 3.75 in. (5.1 x 9.5 cm), Model 44102, Sterile, Hypoallergenic.
Potential contamination of the product with latex adhesive residual.
US
Macherey Nagel Gmbh & Co. Kg recalls Reagent Test Strips for Blood, Urobilinogen, Bilirubin, Protein, Nitrite, Ketones, Ascorbic Acid, Glucose, pH, Specific Gravity and…
The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer by way of a Henry Schein Inc. (HSI) customer complaint.…
FDA · DeviceMacherey Nagel Gmbh & Co. KgUS Nationwide distribution in the states of CO, FL, IN, NV, TX.Official notice →US
Maquet Medical Systems USA recalls CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.
The firm identified that the measured patient leakage current in some devices significantly exceeded expected levels.
FDA · DeviceMaquet Medical Systems USAUS Nationwide distribution in the states of NJ & TX.Official notice →US
Philips North America Llc recalls Ingenia 1.5T with Evolution Upgrade 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a…
Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation
FDA · DevicePhilips North America LlcWorldwide - US Nationwide distribution in the states of CA, KS, MS, NH and the…Official notice →US
Andover Healthcare Inc. recalls Sterile CoFlex NL: Latex Free Flexible Cohesive Bandage, Tan, 4INx5YD (5400S)
A packaging seal gap could impact the sterility of the product.
FDA · DeviceAndover Healthcare Inc.US Nationwide distribution in the states of CA, CO, GA, IA, IN, OH, MN, NC,…Official notice →