CareFusion 303, Inc. recalls Product Name [REF]: Supply Server, CCE Enterprise SW Site License (2-5), BD Pyxis Logistics [806], BD Pyxis MedStation 4000 Main…
During automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then installed this may lead to an error preventing user access to the…
Read the official notice →- Date
- 2025-07-02
- Country
- United States
- Brand
- CareFusion
- Product
- Product Name [REF]: Supply Server, CCE Enterprise SW Site License (2-5), BD Pyxis Logistics [806], BD Pyxis MedStation 4000 Main…
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution including in the states of PR, DE, GU,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More CareFusion recalls
- 2025-09-03
US
CareFusion 303, Inc. recalls BD Pyxis Product Name / UDI-DI code / Catalog No. BD Pyxis" MedStation" 4000 Main / 10885403512629 / 303 BD Pyxis" MedStation" ES Tower /…
Antivirus software was not consistently installed on impacted devices during the implementation process.
FDA · DeviceCareFusion 303, Inc.Worldwide - U.S Nationwide distribution including in the states of AK, AL, AR,…Official notice →- 2025-08-20
US
CareFusion 303, Inc. recalls Description/REF: BD PYXIS MEDBANK MINI 1FH-1FM/169-137, BD PYXIS MEDBANK MINI 4HM/169-113, BD PYXIS MEDBANK MINI 4HH/169-112, BD PYXIS…
Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.
FDA · DeviceCareFusion 303, Inc.US Nationwide distribution including in the states of NY, MA, KY, VA, OR, OH,…Official notice →- 2025-08-13
US
CareFusion 303, Inc. recalls BD Alaris Pump Infusion Sets REF: 10013186, BD Alaris Pump Infusion Set, Back Check Valve, 3 SmartSite Y-sites LVP 60D DEHP FREE 3SS CV;…
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion…
FDA · DeviceCareFusion 303, Inc.US Nationwide Distribution to states of: AK AL AR AS AZ CA CO CT DC DE FL GA…Official notice →- 2025-05-28
US
CareFusion 213, LLC recalls BD ChloraPrep Clear, 2% x/v chlorhexidine gluconate (CHG) and 70% v/v isopropyl alcohol (IPA), Patient Preoperative Skin Preparation,…
Lack of Assurance of Sterility: The affected product may exhibit an open seal on the packaging of the applicator.
- 2025-04-30
US
CareFusion 303, Inc. recalls BD PYXIS MEDBANK (Name, REF): MEDPASS MODULE, 139088-01; MINI 1FH-1FM, 169-137; MINI 1HH-1FH-1HM, 169-101; MINI 1HH-1HM-1FM, 169-102; MINI…
Users trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a software issue that results in 2 destock transactions and 2 stock…
FDA · DeviceCareFusion 303, Inc.US Nationwide distribution including in the states of MO, MI, FL, PA, SC, OR,…Official notice →- 2025-04-23
US
CareFusion 303, Inc. recalls MEDFLEX, REF: 1119-00; MEDBANK MAIN, REF: 1137-00; MEDFLEX 2.0, REF: 1139-00; MEDBANK TOWER MAIN, REF: 1145-00; MEDBANK MINI MAIN, REF:…
If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation…
FDA · DeviceCareFusion 303, Inc.US Nationwide distribution in the states of KS, TX, CA, PA, NJ, FL, IN, AZ,…Official notice →
Other drugs & medical recalls in the United States
- 2025-07-02
US
Limacorporate S.p.A recalls PRIMA 4MM ECCENTRICAL ADAPTOR WITH SCREW, REF:1367.15.704, Sterile
There is the potential that the eccentrical adaptor with screw, black laser marking, use to identify the position may be missing.
FDA · DeviceLimacorporate S.p.AWorldwide - US Nationwide distribution in the states of VA, FL, SC, MS, NY, TX…Official notice →US
Becton Dickinson & Co. recalls BD BACTEC MGIT 960 PZA Kit. IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to…
BD has received additional complaints of intermittent false resistance results for PZA during susceptibility testing of Mycobacterium tuberculosis isolates.
FDA · DeviceBecton Dickinson & Co.Worldwide distribution - US Nationwide and the countries of AR, AU, BD, BE,…Official notice →US
Terumo Cardiovascular Systems Corporation recalls CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial…
Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in…
FDA · DeviceTerumo Cardiovascular Systems CorporationUS Nationwide distribution in the states of California, Delaware, Hawaii,…Official notice →US
Insulet Corporation recalls Omnipod 5 Automated Insulin Delivery System, iOS Application. Model Number: PT-000664/M009-S-AP. Software Version: all Product…
Insulet Corporation is recalling Omnipod 5 iOS App, due to a software design issue in which a comma is not recognized as a decimal separator and the user cannot enter a decimal separator in any manual entry screen if…
FDA · DeviceInsulet CorporationUS market only- Although there are 122823 downloads of the application, our…Official notice →US
RECORDATI RARE DISEASES INC. recalls Caphosol Artificial Saliva (32 doses sachet box). 64 x 15 mL sachets aqueous solution (32 sachets Solution A Phosphate, 32 sachets…
Out of specification stability test result for the Caphosol B solution from process validation lots
FDA · DeviceRECORDATI RARE DISEASES INC.Worldwide - US Nationwide distribution in the states of GA, OH, TN, TX, VA and…Official notice →US
Encore Medical, LP recalls Brand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm Model/Catalog Number: 506-04-126 Software Version:…
Package labeling for Altivate Reverse Torx Peripheral screw contains a different size then actual included product.
FDA · DeviceEncore Medical, LPAt time of quarantine: RFK Medical Inc. Virginia (VA) CE102 506-04-122…Official notice →