CareFusion 303, Inc. recalls (1) BD Pyxis MedStation ES, Catalog No. 323; (2) BD Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) BD Pyxis MedStation ES Tower,…
Labeling is insufficient regarding delays in access to medication. The firm is strengthening the label as an additional risk mitigation.
Read the official notice →- Date
- 2025-02-19
- Country
- United States
- Brand
- CareFusion
- Product
- (1) BD Pyxis MedStation ES, Catalog No. 323; (2) BD Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) BD Pyxis MedStation ES Tower,…
- Classification
- Class II
- Units
- Domestic distribution nationwide US. International distribution to North…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More CareFusion recalls
- 2025-03-05
US
CareFusion 303, Inc. recalls BD Pyxis MedFlex 2.0 REF 1139-00 Bio-ID Spare Parts REF 353200-01 The BD Pyxis" MedFlex is an Automated Dispensing Cabinet (ADC) intended…
Due to fingerprint scanner failing resulting in the scanner heating up.
FDA · DeviceCareFusion 303, Inc.Worldwide - U.S. Nationwide distribution including in the states of AL, AZ,…Official notice →US
CareFusion 303, Inc. recalls BD Pyxis: MedFlex, REF: 1119-00; MedBank Tower REF: 1145-00; MedBank MedPass, REF: 139088-01 and associated BD Pyxis user/software/safety…
Automated dispensing device labeling is being updated to strengthen labeling around manual access to medications to address the risk of potential delays in retrieving contents/ accessing medications that may result…
FDA · DeviceCareFusion 303, Inc.US Nationwide distribution including in the states of AZ, FL, NJ, KY, GA, CA,…Official notice →- 2025-02-19
US
CareFusion 303, Inc. recalls BD Pyxis MedStation 4000 REF: 303. Intended to securely store and dispense medications to a qualified and authorized healthcare provider.
Retroactive-Due to increase in complaints, their is a potential for drawer and/or door failures of the medication cabinet which could result in delay in accessing medications.
FDA · DeviceCareFusion 303, Inc.US: Nationwide OUS:JP IT IN DE GB KR NL TW ES BE AU BH CA BR NZ FR AR QA MX CL…Official notice →US
CareFusion 303, Inc. recalls (1) Pyxis MedStation ES, Catalog No. 323; (2) Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) Pyxis MedStation ES Tower, Catalog No.…
Software issues could potentially result in: 1) delays in accessing dispensing software application, 2) wrong therapy (dose, item, timing, etc.), or 3) unauthorized access and/or compliance related issues
FDA · DeviceCareFusion 303, Inc.Domestic distribution nationwide. International distribution to North America,…Official notice →- 2025-02-05
US
CareFusion 303, Inc. recalls BD Pyxis MedStation ES AUX Tower REF 343 BD Pyxis " MedStation 4000 AUX Tower REF 314 The BD Pyxis" MedStation ES and Pyxis" MedStation…
Due to an increase in complaints of door latch assembly failures that may cause the doors to be left in an open state potentially causing inventory discrepancies, a delay in access to medication and/or diversion of…
FDA · DeviceCareFusion 303, Inc.U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS,…Official notice →- 2025-01-29
US
CareFusion 303, Inc. recalls BD Pyxis Med ES Auxiliary (AUX), REF 343
Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure,…
Other drugs & medical recalls in the United States
- 2025-02-19
US
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) KIT,NASAL SEPTOPLASTY,FAIRVIEW, Catalog Number:PE29NSFG4; 2) KIT,NASAL SEPTOPLASTY,FAIRVIEW, Catalog…
Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.
FDA · DeviceCardinal Health 200, LLCWorldwide distribution - US Nationwide and the countries of Canada, Saudi…Official notice →US
CareFusion 303, Inc. recalls BD Pyxis MedStation 4000 REF: 303. Intended to securely store and dispense medications to a qualified and authorized healthcare provider.
Retroactive-Due to increase in complaints, their is a potential for drawer and/or door failures of the medication cabinet which could result in delay in accessing medications.
FDA · DeviceCareFusion 303, Inc.US: Nationwide OUS:JP IT IN DE GB KR NL TW ES BE AU BH CA BR NZ FR AR QA MX CL…Official notice →US
MICROVENTION INC. recalls MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile R, CE0297. The intended…
Potential for Polytetrafluoethylene (PTFE) material encasing the stent loop may prevent the stent from being pushed out of the introducer.
US
CareFusion 303, Inc. recalls (1) Pyxis MedStation ES, Catalog No. 323; (2) Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) Pyxis MedStation ES Tower, Catalog No.…
Software issues could potentially result in: 1) delays in accessing dispensing software application, 2) wrong therapy (dose, item, timing, etc.), or 3) unauthorized access and/or compliance related issues
FDA · DeviceCareFusion 303, Inc.Domestic distribution nationwide. International distribution to North America,…Official notice →US
Max Mobility LLC recalls Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DC use in conjunction with the SmartDrive MX2+ Wheelchair Power Assist…
Due to faulty circuit board in speed control remote use in conjunction with a wheelchair power assist, their is a potential for lost of control of device potentially resulting in minor and major injuries.
FDA · DeviceMax Mobility LLCUS: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO…Official notice →US
Ithera Medical Gmbh recalls The MSOT Acuity Echo is a Class 4 medical laser device. The MSOT Acuity is a mobile assembly of electrically-powered components intended…
The MSOT Acuity Echo does not include an adequate instruction for safe operation of the device. The previous version of the user manuals contained a wrong value for the Nominal Ocular Hazard Distance (NOHD).
FDA · DeviceIthera Medical GmbhWorldwide distribution: US (Nationwide) and OUS (foreign) to countries of:…Official notice →