Boston Scientific Corporation recalls Torflex Transseptal Guiding Sheath Kit (TFK-M) TF85-32-63-45 TORFLEX 8.5F 63CM 45DEG TF85-32-63-55 TORFLEX 8.5F 63CM 55DEG Used…
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may…
Read the official notice →- Date
- 2025-07-16
- Country
- United States
- Brand
- Boston Scientific
- Product
- Torflex Transseptal Guiding Sheath Kit (TFK-M) TF85-32-63-45 TORFLEX 8.5F 63CM 45DEG TF85-32-63-55 TORFLEX 8.5F 63CM 55DEG Used for the…
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Boston Scientific recalls
- 2025-08-27
US
Boston Scientific Corporation recalls ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282,…
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or…
FDA · DeviceBoston Scientific CorporationWorldwide Distribution. US nationwide. International distribution worldwide.Official notice →- 2025-08-20
US
Boston Scientific Corporation recalls FARAWAVE 1.0 Pulsed Field Ablation Catheter, Product ID M004PF41M401
The potentially impacted units were manufactured using specific equipment that may have caused cracks in the electrode bands on the catheter's distal end.
US
Boston Scientific Corporation recalls Carotid WALLSTENT Monorail Endoprosthesis Closed Cell Self-Expanding Stent, stent and stent delivery system. Material numbers:…
Certain batches are being removed due to an increase in complaints received in which physicians encountered greater than anticipated resistance while attempting to withdraw the stent delivery system (SDS) from the…
FDA · DeviceBoston Scientific CorporationDomestic US Nationwide distribution including Puerto Rico. International…Official notice →- 2025-05-07
US
Boston Scientific Corporation recalls IceSeed 1.5 CX S NEEDLE OUS, Cryoablation Needle, REF H7493967334100. The needles are intended to convert high-pressure gas to…
Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon…
- 2025-02-05
US
Boston Scientific Corporation recalls AXIOS Stent and Delivery System 10mmx10mm UPN: M00553740
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
FDA · DeviceBoston Scientific CorporationNationwide including Puerto Rico Foreign: To be providedOfficial notice →- 2025-01-22
US
Boston Scientific Corporation recalls PROPONENT DR Pacemaker, SL (Model Number L201), (VDD) SL (Model Number L209), SL MRI (Model Number L211), EL (Model Number L221),…
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to…
FDA · DeviceBoston Scientific CorporationUS (Nationwide) including Puerto Rico and US Virgin Islands and OUS (foreign)…Official notice →
Other drugs & medical recalls in the United States
- 2025-07-23
US
SUN PHARMACEUTICAL INDUSTRIES INC recalls Lisdexamfetamine Dimesylate Capsules, 60 mg, CII, 100-count bottles, Rx only, Manufactured by: OHM Laboratories, New…
Failed Dissolution Specifications: An out of specification results observed in dissolution test during analysis at 9 month long term stability station (25¿C, 60%RH).
FDA · DrugSUN PHARMACEUTICAL INDUSTRIES INCRecalled units were distributed to 17 distributors who may have further…Official notice →- 2025-07-16
US
Alcon Research LLC recalls Brand Name: Alcon Laboratories, Inc. Product Name: Constellation Vision System Vitrectomy Procedure Pak Model/Catalog Number: Many - See…
Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable,…
FDA · DeviceAlcon Research LLCWorldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →US
Preat Corp recalls Preat, Astra EV-compatible 4.2mm Digital Analog REF: 9006603 (single) and 9006603-10(10 pack), Rx Only
Due to manufacturing error, digital analog rotational feature is 180 degrees off.
FDA · DevicePreat CorpUS Nationwide distribution in the states of AL, CA, FL, GA, KY, MD, MN, NY,…Official notice →US
Medtronic Navigation, Inc.-Boxborough recalls O-arm O2 Imaging System. Mobile X-Ray System.
A mechanical component that supports the O2 gantry to the support structure in affected devices may have been damaged during manufacture.
FDA · DeviceMedtronic Navigation, Inc.-BoxboroughUS Nationwide distribution in the states of IN, MN, MS.Official notice →US
Neurovision Medical Products Inc recalls Brand Name: Hummingbird Product Name: Stimulated EMG Kit with Hummingbird - Model/Catalog Number: REF I-D-BMF90, 1 Single-Use…
Product packaging contains the wrong part number and kit contents.
US
Canary Medical, Inc. recalls Brand Name: Canturio¿ Smart Extension Product Name: CSE Implant Model/Catalog Number: 43-5570-030-14 Product Description: Tibial stem…
Due to labeling error. Product incorrectly labeled.