Boston Scientific Corporation recalls ROTAWIRE Drive and wireClip Torquer, ROTAWIRE Drive Floppy
Pouch seals may be open, compromising sterility.
Read the official notice →- Date
- 2022-04-20
- Country
- United States
- Brand
- Boston Scientific
- Product
- ROTAWIRE Drive and wireClip Torquer, ROTAWIRE Drive Floppy
- Classification
- Class II
- Units
- Distribution in the US and territories, Japan, and Canada
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Boston Scientific recalls
- 2022-07-27
US
Boston Scientific Corporation recalls GreenLight HPSEA Laser Fibers, UPN 0010-2092
Several complaints were received for the affected lot; users received a "Fiber type does not match card type" error message on the GreenLight XPS Laser console, necessitating the replacement of the laser fiber and…
FDA · DeviceBoston Scientific CorporationDomestic distribution to AR CA IL ME MI NC TX Foreign distribution to…Official notice →- 2022-05-25
US
Boston Scientific Corporation recalls EXALT Model D Single- Use Duodenoscope (Box 2) UPN: M00542421 Intended Use: sterile, single-use, flexible video duodenoscope…
Updated Instructions for Use (IFU) for EXALT Model D Single-Use Duodenoscope highlight the known risk of perforation and provide best clinical practices related to insertion, advancement, and removal of this device.
FDA · DeviceBoston Scientific CorporationWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2022-05-04
US
Boston Scientific Corporation recalls SpaceOAR System, UPN SO-1010, SO-2101, SO-3101, and SO-4101. Used to temporarily position the anterior rectal wall away from the…
Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct…
- 2022-03-23
US
Boston Scientific Corporation recalls HYDRATOME RX 44-20MM/450CM Material Number: M00583060
Sterility of device is compromised due to a sterile barrier breach
FDA · DeviceBoston Scientific CorporationWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2021-12-22
US
Boston Scientific Corporation recalls Boston Scientific ROTAPRO, Rotational Atherectomy System Console, Model numbers H74939309010 (US) and H74939309030 (EU).
Manufacturing instructions did not include the directions to not retest the pneumatic kit component of the ROTAPRO console upon failure, an internal nonconformity.
FDA · DeviceBoston Scientific CorporationWorldwide distribution - US Nationwide distribution in the states of FL, NV,…Official notice →- 2021-12-08
US
Boston Scientific Corporation recalls HURRICANE RX DILATION BALLOON 6MM 4CM- -for endoscopic dilatation of strictures of the biliary tree and the Sphincter of Oddi. UPN:…
Balloon pinhole results in losss pressure or fails to gain or maintain pressure, causing prolonged procedure
Other drugs & medical recalls in the United States
- 2022-04-20
US
GE Healthcare, LLC recalls MUSE Cardiology Information System, Model Numbers: a) 3101011-040, b) 3101010-401, c) 3101010-009, d) 3101010-008
Two scenarios may cause edits to measurements and diagnosis statements can be lost after a test is Signed in the MUSE NX web client.
FDA · DeviceGE Healthcare, LLCUS and Australia, Austria, Bahrain, Belgium, Brazil, Canada, Denmark, England,…Official notice →US
Sensory Medical Inc recalls Cub 2 Enclosed Bed Canopy System
Due potential misuse of the product, firm is updating warnings and precautions booklet for products distributed before December 16, 2021.
FDA · DeviceSensory Medical IncU.S. Nationwide distribution to states including: AK, AZ, CA, CO, CT, DE, FL,…Official notice →US
Volcano Corp recalls MM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Numbers: 300002018101, 300002018102, used with IntraSight…
Multi-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged, which may affect their usability. They could become non-functional. Therefore, bottom protective beams will…
FDA · DeviceVolcano CorpUS: AZ, IL, NV, MI, FL, NC, AK, PA, WI, MN, CA, AR, NJ, TX, OR, NM, SD, CO,…Official notice →US
Nextremity Solutions recalls Nextra Hammertoe Correction System 3.5 Middle and 3.2 Proximal Saw Cut SKU: NX-3532K-SC Orthopedic surgical instrument
Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery
FDA · DeviceNextremity SolutionsDistribution US Nationwide, Japan, and the NetherlandsOfficial notice →US
Integra LifeSciences Corp. recalls Integra Universal Flexible Arm part number REF 1362275
Excess force used to tighten the Universal Flexible Arm can cause the internal cable to break at the tip, resulting in an unusable condition.
FDA · DeviceIntegra LifeSciences Corp.US Nationwide Distribution: AL, AZ, FL, GA, IL, MA, MD, NY, OH, OR, PA, TXOfficial notice →US
Stanbio Laboratory, LP recalls Separation Technology Creamatocrit Plus, Catalog Number: 100-146
The device was being marketed as an IVD without a 510(k). Therefore, performance characteristics of these breast milk analyzers have not been adequately validated. An erroneous result could lead to sub-optimal…
FDA · DeviceStanbio Laboratory, LPUS Nationwide Distribution CA, TX, OK, SD, IL, NV, MO, FL, WA, KY, WI, PA, NC,…Official notice →