Blue Belt Technologies, Inc recalls CORI Starter Kit, Real Intelligence CORI, CORI STARTER KIT - UNIVERSAL.Intended to assist the surgeon in providing software-defined…
CORI Surgical System with Real Intelligence Software exhibits an increased Marker Registration Error (MRE) which may cause tracker arrays to flicker on screen of the CORI Surgical System during use, and may cause…
Read the official notice →- Date
- 2024-06-19
- Country
- United States
- Brand
- Blue Belt Technologies
- Product
- CORI Starter Kit, Real Intelligence CORI, CORI STARTER KIT - UNIVERSAL.Intended to assist the surgeon in providing software-defined…
- Classification
- Class II
- Units
- Nationwide Foreign: Australia, Austria, Belgium, Brazil, Canada, Chile, China,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Blue Belt Technologies recalls
- 2024-01-17
US
Blue Belt Technologies, Inc recalls CORI REAL INTELLIGENCE Robotic Drill, Product No. ROB10013 and ROB10013S
Possibility that the user is unable to resolve drill disconnection error messages.
- 2023-10-11
US
Blue Belt Technologies, Inc recalls Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299
CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational…
Other drugs & medical recalls in the United States
- 2024-06-19
US
Fresenius Medical Care Holdings, Inc. recalls US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers:…
Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is…
FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AL, IN, AK, WI.Official notice →US
Aesculap Inc recalls Aeos Robotic Digital Microscope, Product Code: PV010
Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.
FDA · DeviceAesculap IncUS Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE,…Official notice →US
Integra LifeSciences Corp. recalls CODMAN BACTISEAL EVD Catheter Set - Indicated for gaining access to the ventricles of the brain and can be used with dimensionally…
Defect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal that may compromise the product sterility.
FDA · DeviceIntegra LifeSciences Corp.Worldwide - US Nationwide distribution in the state of TX and the countries of…Official notice →US
B Braun Medical Inc recalls Stimuplex A, 30 DEG, 24GX1", 0.55x25mm, Catalogue Number: 4894251.
The devices are labeled "not made with DEHP" or the DEHP free symbol; however, the glue used contains traces of DEHP.
FDA · DeviceB Braun Medical IncWorldwide distribution - US Nationwide and the country of Canada.Official notice →US
Rygel Advanced Machines d/b/a Omtech Laser recalls OMTech Desktop Laser Engraver
Lack of redundant interlocks, missing required labels and certification labels as required by 21 1010.2.
US
Canon Medical System, USA, INC. recalls MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage…
The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of…
FDA · DeviceCanon Medical System, USA, INC.US: PA, SC, IA, AL, MI, AR, SD, TX, ID, KY, NY, MN, MD, OH, WI, NV, CT, AZ,…Official notice →