Belmont Instrument Corporation recalls CritiCool, thermoregulation device
Premature pump failure may lead to the device displaying a HALT 4 error, from which the device cannot recover even after device reboot. The issue may potentially lead to delayed treatment and a failure to deliver the…
Read the official notice →- Date
- 2022-11-09
- Country
- United States
- Brand
- Belmont Instrument
- Product
- CritiCool, thermoregulation device
- Classification
- Class II
- Units
- Domestic distribution nationwide. Foreign distribution worldwide.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Belmont Instrument recalls
- 2025-06-18
US
Belmont Instrument LLC recalls Allon 2001. Part Number: 200-00263
An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the…
FDA · DeviceBelmont Instrument LLCWorldwide - US Nationwide distribution in the states of CA, DC, LA, MA, NM,…Official notice →- 2025-05-07
US
Belmont Instrument LLC recalls 3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume Reservoirs (P/N 903-00018)…
Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser.
FDA · DeviceBelmont Instrument LLCNationwide distribution to AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, LA, MA,…Official notice →US
Belmont Instrument LLC recalls Hyperthermia Pump Procedure Kit. Model Number: 902-00045. Sterile, single-use, disposable set used with the Hyperthermia Pump. The…
Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.
FDA · DeviceBelmont Instrument LLCWorldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC,…Official notice →- 2025-02-12
US
Belmont Instrument LLC recalls Belmont Medical 3-Spike Disposable Set used with the Belmont Rapid Infuser RI-2. Designed to be used in general operation in hospital or…
Disposable set may leak during priming from a female quick connector due to a crack in the female quick connector, may lead to a potential delay in treatment since the set will need to be replaced by another set
FDA · DeviceBelmont Instrument LLCWorldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →
Other drugs & medical recalls in the United States
- 2022-11-09
US
Johnson & Johnson Surgical Vision, Inc. recalls THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac…
Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation…
FDA · DeviceJohnson & Johnson Surgical Vision, Inc.Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO,…Official notice →US
Rebiscan, Inc. recalls Blinq Pediatric Vision Scanner-Pediatric Vision Scanner-intended for automated detection of misalignment of the visual axes using…
Unable to charge the battery pack- battery packs will allow the batteries to be charged if the voltage of any cell is below 2.5V. When this occurs, the battery pack will not accept the charge, could vent and possibly…
FDA · DeviceRebiscan, Inc.Worldwide distribution - US Nationwide and the countries of New Zealand,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls CENTURION MEDICAL PRODUCTS NICU PICC INSERTION BUNDLE W/O, REF CVI4470. Convenience kit used in medical procedures.
Medline received complaints on some of the breather pouches with in the sterile surgical kits show openings in the seal.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution in the states of CA, MN, NC, NY, OK, PA, TN, TX.Official notice →US
Boston Scientific Corporation recalls EMBLEM S-ICD, Subcutaneous Implantable Defibrillators, Model A209, sterile.
There is an incorrect manufacturing date/timestamp within the software which causes an inaccurate display of battery capacity.
FDA · DeviceBoston Scientific CorporationWorldwide distribution - US Nationwide distribution in the countries of AL,…Official notice →US
Ethicon, Inc. recalls STRATAFIX SPIRAL PGA-PCL UNI UD 20CM USP2-0(M3) S/A SH. indicated for use in soft tissue approximation where the use of absorbable sutures…
Packages may contain a PS-1 cutting edge needle and not the intended SH taper point needle may result in more tissue trauma than is anticipated for taper point needles, which may necessitate repair
US
Boston Scientific Corporation recalls EKOS Control Unit 4.0 console, a component of the EKOS Control System 4.0, which is a component of the EkoSonic Endovascular System.
Consoles were shipped without an instruction for use (IFU) or with the incorrect IFU.
FDA · DeviceBoston Scientific CorporationDistribution was made to CA, IL, KY, OH, OK, RI, TX. There was no…Official notice →