Drugs & medicalClass II

Becton Dickinson & Co. recalls BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species. Catalog (Ref) Number: 221267

Boxes contain a reduced concentration of Nicotinamide Adenine Dinucleotide and does not provide enough of the blood factors necessary to support Haemophilus growth. The reduction could cause a false negative result for…

Read the official notice →
Date
2022-12-21
Country
United States
Brand
Becton Dickinson &
Product
BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species. Catalog (Ref) Number: 221267
Classification
Class II
Units
Worldwide distribution - US Nationwide and the countries of Canada, India,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Becton Dickinson & recalls

  1. 2023-06-14
  2. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Veritor System for Rapid Detection of SARS-CoV-2 & Flu A+B, 30 Single Use Test Devices- Intended for the in vitro, simultaneous…

    IFU- Instructions for Use (IFU) package insert and both the IFU packaged with the test kit and the electronic version on the BD website not approved by FDA,specifically describing specimen type, could cause use of the…

    FDA · DeviceBecton Dickinson & Co.Worldwide distribution - US Nationwide and the countries of CA, CL, HK, TW.Official notice →
  3. 2023-04-26
  4. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Kiestra InoqulA+ with BD Kiestra" InoqulA+" and BeA Software version 5.1 and 5.1.1 as part of the Icefall A platform. An in vitro…

    Upon installation of BD Kiestra" InoqulA+" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, plate information was not visible in Synapsys" after processing has occurred. This could possibly lead to a…

    FDA · DeviceBecton Dickinson & Co.Worldwide - US Nationwide distribution in the states of IL, NY and the…Official notice →
  5. 2023-03-22
  6. US
    Drugs & medicalClass II

    Becton Dickinson & Company recalls BD Ultra-Fine Insulin Syringe 0.5mL, 8mm, 31G-Insulin Syringes are designed for administering insulin in the treatment of diabetes…

    Nonsterile syringe potentially exposing the patient to pathogenic microorganisms and a risk of local infection.

    FDA · DeviceBecton Dickinson & CompanyWorldwide distribution - US Nationwide and the countries of Chile, Colombia,…Official notice →
  7. 2022-11-23
  8. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibility of most aerobic and facultative…

    Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false…

    FDA · DeviceBecton Dickinson & Co.US Nationwide distribution.Official notice →
  9. 2022-10-26
  10. US
    Drugs & medicalClass II

    Becton Dickinson & Company recalls BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg, Catalog Number 367001. For collecting, transporting,…

    The affected product contains isobutylene which has recently demonstrated potential interference in determination of methanol by gas chromatography methods. The interference may lead to false positive results in…

    FDA · DeviceBecton Dickinson & CompanyDistribution US nationwide, Canada, New Zealand, and Hong Kong.Official notice →
  11. 2022-08-24
  12. US
    Drugs & medicalClass II

    Becton Dickinson & Company recalls BD PhaSealTM Y-Site Connector (C80)- airtight and leak-proof Closed System Drug Transfer Device (CSTD) that mechanically prohibits the…

    Incorrect expiration date listed on the product labels. The expiration date listed on the product label is beyond the expiration date of the product.

    FDA · DeviceBecton Dickinson & CompanyUS Nationwide Distribution CA FL GA IL MD MI OR PA PR TN TXOfficial notice →
All 36 Becton Dickinson & recalls →

Other drugs & medical recalls in the United States

  1. 2022-12-21
  2. US
    Drugs & medicalClass II

    Allison Medical, Inc. recalls CarePoint 60 mL Syringe Without Needle, REF: 04-7560

    Syringe without needle may have a green substance on the plunger. Substance found to be nontoxic but its effect on drug potency or interaction is unknown.

    FDA · DeviceAllison Medical, Inc.US: IA, LA, MS, FL, PR; OUS: BARBADOSOfficial notice →
  3. US
    Drugs & medicalClass II

    Leica Microsystems, Inc. recalls Leica MICROSYSTEMS M530 OHX, Part Number 10448737, surgical microscope system

    There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific…

    FDA · DeviceLeica Microsystems, Inc.US NationwideOfficial notice →
  4. US
    Drugs & medicalClass II

    ConMed Corporation recalls Foot Controlled Electrosurgical Suction Coagulator, Electrosurgical Suction Coagulator 10FR, Catalog Number 130187

    Suction ports may potentially be occluded on the affected devices. If this occurs, the affected devices will not be able to perform the suction function during use, which could cause a delay in procedure and could…

    FDA · DeviceConMed CorporationDomestic distribution nationwide. Foreign distribution to Colombia, Ecuador,…Official notice →
  5. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls ARTISTE, ONCOR and PRIMUS with syngo RT Therapist with software version 4.3.1 MR3. Material Number: 8162815 Intended Use:…

    When selecting the site fraction group for treatment, the User may select the wrong site for treatment, and potentially deliver the dose to wrong isocenter

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution in the states of CA, IA, LA, NE, NV, OH, SC.Official notice →
  6. US
    Drugs & medicalClass II

    Zimmer, Inc. recalls NexGen Option Stemmed Tibial Component Size 7, Item Number 00-5986-057-01

    Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial…

    FDA · DeviceZimmer, Inc.Worldwide - US Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass I

    ARROW INTERNATIONAL Inc. recalls Arrow AutoCAT2 Intra-Aortic Balloon Pump, AEROAUTOCAT2 WAVE REFURBISHED, REF IAP-0535X (IPN000329), cardiac pump

    There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.

    FDA · DeviceARROW INTERNATIONAL Inc.WorldwideOfficial notice →
See all drugs & medical recalls →