Drugs & medicalClass II

Baxter Healthcare Corporation recalls XScribe CP Cardiac Stress Testing System, version 6, with below product names and codes: a) X-SCRIBE CPI KIT WIRELESS USB 608 MHz,…

Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.

Read the official notice →
Date
2024-01-17
Country
United States
Brand
Baxter Healthcare
Product
XScribe CP Cardiac Stress Testing System, version 6, with below product names and codes: a) X-SCRIBE CPI KIT WIRELESS USB 608 MHz,…
Classification
Class II
Units
United States (including Puerto Rico), Canada, Australia, Bangladesh, Brunei,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Baxter Healthcare recalls

  1. 2024-03-20
  2. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls STERIS Boot Assembly, Traction Device, OT 1000, Part Number REF 10009192

    Three lots of the Boot Assembly, Traction Device are being recalled due to due to a customer report of an incorrectly oriented locking pin which caused the boot to detach from the traction system. If an impacted device…

    FDA · DeviceBaxter Healthcare CorporationUS Nationwide distribution in the state of Alabama.Official notice →
  3. 2024-02-28
  4. US
    Drugs & medicalClass I

    Baxter Healthcare Corporation recalls Baxter Exactamix Pro 1200, REF EXM12DY

    An error was identified in software versions 2.0.8 and 2.1.8 while using the "Use Some Overfill" feature which may lead to over-delivery of an ingredient. This issue can result in redundant ingredient delivery of the…

    FDA · DeviceBaxter Healthcare CorporationUS: AL, AZ, CA, FL, GA, IA, KY, MD, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OR,…Official notice →
  5. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Norepinephrine Bitartrate in 5% Dextrose Injection, 8 mg/ 250 mL (32 mcg/mL), For Intravenous Infusion Only, 250 mL vial, Rx Only,…

    Incorrect product concentration on the overwrap label: The overwrap label incorrectly identified the product strength as 4 mg / 250 mL; however, the primary bag label correctly identified the product strength as 8 mg /…

    FDA · DrugBaxter Healthcare CorporationNationwide in the USAOfficial notice →
  6. 2024-01-17
  7. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Baxter Flo-Thru Intraluminal Shunt, Product Codes: a) Product Code FT12100, 1.00 GLOBAL; b) Product Code FT12125, 1.25 GLOBAL; c)…

    Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or…

    FDA · DeviceBaxter Healthcare CorporationUS Nationwide distribution.Official notice →
  8. 2023-12-20
  9. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Baxter Amia Automated Peritoneal Dialysis (APD) Cycler Set, REF 5C5479

    There is to an increase in complaints related to failure alarms for Wet Cassette Integrity Test occurring on the Amia devices.

    FDA · DeviceBaxter Healthcare CorporationWorldwide distribution - US Nationwide and the country of Canada.Official notice →
  10. 2023-12-13
  11. US
All 123 Baxter Healthcare recalls →

Other drugs & medical recalls in the United States

  1. 2024-01-17
  2. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls QMS Tacrolimus Calibrators, Product code 10015573

    Ortho-Clinical Diagnostics, Inc. is recalling QMS Tacrolimus Calibrators, Lot 74859619 expiry date: May 11, 2024, due to improper storage that is expected to degrade the quality of the calibrator and render it…

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Domestic distribution to IL, IN, FL, NJ, NM, TX. International distribution to…Official notice →
  3. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Medline Gown Poly W/Breath Sleeves XL AAMIK Level 4 (Eclipse gown), REF SPT-2208CS, 100/pack, Non-Sterile.

    Surgical gowns were manufactured with the wrong sleeve.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldDistribution was made to MD, OH, TN, TX, and WI. There was no…Official notice →
  4. US
    Drugs & medicalClass II

    Blue Belt Technologies, Inc recalls CORI REAL INTELLIGENCE Robotic Drill, Product No. ROB10013 and ROB10013S

    Possibility that the user is unable to resolve drill disconnection error messages.

    FDA · DeviceBlue Belt Technologies, IncDomestic: Nationwide Distribution.Official notice →
  5. US
    Drugs & medicalClass II

    BioFire Diagnostics, LLC recalls The BIOFIRE Respiratory Panel 2.1 (RP2.1) is a PCR-based multiplexed nucleic acid test intended for use with the BIOFIRE FILMARRAY 2.0…

    A manufacturing event may contribute to elevated rates of false negative results while using gastrointestinal panel and respiratory panel, which could lead to inappropriate antibiotic treatment, no treatment, delay of…

    FDA · DeviceBioFire Diagnostics, LLCUS: NC, MO, NE, CT, IL, TN, NJ, OH, MA, CA, FL, UT, IA, KS, TX, MDOfficial notice →
  6. US
    Drugs & medicalClass II

    Immunostics, Inc. recalls Detector H. Pylori, Catalog Number IHP-30

    Labeling for affected lots incorrectly indicates the expiration date is 10/31/2023. However, the correct expiration date is 10/31/2022. Use of the product after its expiration may cause false positive, false negative,…

    FDA · DeviceImmunostics, Inc.Product distributed to CA and MO.Official notice →
  7. US
    Drugs & medicalClass II

    Icotec Ag recalls VADER pedicle system, torque wrench, Catalog Number 42-702

    icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the…

    FDA · DeviceIcotec AgUS: AZ. MI, OH, CA, CO, TX, NY, NC, DE, PAOfficial notice →
See all drugs & medical recalls →