Baxter Healthcare Corporation recalls Progressa Bed Surfaces, intended to be used to treat or prevent pulmonary or other complications associated with immobility,…
The air bladders inside the mattress may move out of position when the head of the bed is elevated, causing a dip in the mattress.
Read the official notice →- Date
- 2025-04-02
- Country
- United States
- Brand
- Baxter Healthcare
- Product
- Progressa Bed Surfaces, intended to be used to treat or prevent pulmonary or other complications associated with immobility, Product Codes…
- Classification
- Class II
- Units
- US Nationwide. Global Distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Baxter Healthcare recalls
- 2025-05-21
US
Baxter Healthcare Corporation recalls Welch Allyn Life2000 Compressor, Product Codes: 1) BT-80-0004; 2) BT-80-0004A; 3) BT-80-0008; 4) BT-80-0008A; 5) RMS010093CP.
A cybersecurity vulnerability was discovered through internal testing.
- 2025-04-23
US
Baxter Healthcare Corporation recalls Welch Allyn Connex Vital Signs Monitor, Product Code/Part Numbers: 1) ) Product Code/Part # 67MXTP-B-ECG3A; 2) ) Product Code/Part…
Undeclared natural rubber band within the packaging of certain configurations of reusable blood pressure cuffs.
US
Baxter Healthcare Corporation recalls Welch Allyn Connex Vital Signs Monitor (CVSM):
There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
- 2025-04-02
US
Baxter Healthcare Corporation recalls Baxter NovumIQ Syringe INFUSION SYSTEM , REF 40800BAXUS
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Novum IQ syringe pumps listed below due to an incorrectly installed gasket. A gasket failure could lead to potential fluid ingress due to…
FDA · DeviceBaxter Healthcare CorporationUS Nationwide distribution in the states of AR, CA, FL, IA, IN, LA, MO, MT,…Official notice →US
Baxter Healthcare Corporation recalls Baxter TruSystem 7500 Hybrid Plus (FC), Product Code 1854087
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
- 2025-03-12
US
Baxter Healthcare Corporation recalls Baxter Sigma Spectrum Infusion System (V6 Platform), Product Code 35700BAX
There is the potential for missing motor mounting screws, which may have occurred during the servicing process.
Other drugs & medical recalls in the United States
- 2025-04-09
US
BIOCON PHARMA INC recalls Atorvastatin Calcium Tablets USP 40 mg, 1000-count bottle, Rx Only, Manufactured for: Biocon Pharma Inc., Iselin, NJ 08830-3009 USA,…
Failed dissolution specifications: lower than specifications
US
Imprimis NJOF, LLC recalls Tropicamide - Proparacaine - Phenylephrine - Ketorolac Sterile Ophthalmic Solution, 1% 0.5% 2.5% 0.5%, 5 mL fill in 15 mL bottle, Imprimis…
Sub-Potent Drug: The stability batch showed a subpotent result, which does not meet the product specifications of 90.0-110.0% of the label claim.
US
The Harvard Drug Group LLC dba Major Pharmaceuticals and… recalls Tolterodine Tartrate Extended-Release Capsules, 2 mg, 30 Capsules (3 x 10) Unit Dose carton, Rx only,…
Failed Dissolution Specifications: Out of specification results obtained during routine stability testing for dissolution.
FDA · DrugThe Harvard Drug Group LLC dba Major Pharmaceuticals and…Nationwide in the USAOfficial notice →- 2025-04-02
US
Checkpoint Surgical Inc recalls Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Small Model/Catalog Number:…
Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to…
US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline ReNewal Reprocessed Livewire Steerable Electrophysiology Catheters, Medline Item Numbers: 401582R, 401904R, 401904R,…
Devices may have higher than expected amounts of bacterial endotoxin.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution in the states of AR, AZ, CA, CO, IA, IL, KS, OR,…Official notice →US
CooperVision, Inc. recalls Brand Names: MyDay Toric, LensCrafters One Day Premium Daily Disposable Contact Lenses For Astigmatism, Pearle Vision 1 Day Premium For…
a limited number of lots were manufactured with an incorrect cylinder power.
FDA · DeviceCooperVision, Inc.Domestic: CA, CO, FL, IL, KY, MI, NC, NE, NJ, NY, OH, OR, UT, VA, WA.…Official notice →