Drugs & medicalClass II

Baxter Healthcare Corporation recalls HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P7501A000033, P7501A000052,…

Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator through the transport handle during transport, resulting in a mild…

Read the official notice →
Date
2023-11-08
Country
United States
Brand
Baxter Healthcare
Product
HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P7501A000033, P7501A000052, P7501A000053,…
Classification
Class II
Units
US Nationwide Distribution in the states of: CA, FL, IN, MA, NY, OH, TX, VA
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Baxter Healthcare recalls

  1. 2023-12-20
  2. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Baxter Amia Automated Peritoneal Dialysis (APD) Cycler Set, REF 5C5479

    There is to an increase in complaints related to failure alarms for Wet Cassette Integrity Test occurring on the Amia devices.

    FDA · DeviceBaxter Healthcare CorporationWorldwide distribution - US Nationwide and the country of Canada.Official notice →
  3. 2023-12-13
  4. US
  5. 2023-11-15
  6. US
    Drugs & medicalClass I

    Baxter Healthcare Corporation recalls Novum IQ Syringe infusion system, Product Code 40800BAXUS

    Baxter is issuing an Urgent Medical Device Correction for the Novum IQ Syringe Pump. Baxter identified that after multiple downstream occlusion alarms, the pump may display an Infusion Complete alarm even though…

    FDA · DeviceBaxter Healthcare CorporationUS Nationwide Distribution to states of: AR, AZ, CA, FL, GA, HI, IA, IN, KS,…Official notice →
  7. 2023-10-25
  8. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Doxil (doxorubicin hydrochloride liposome injection) 50 mg in 25 mL (2 mg/mL), Single-Dose Vial, Rx only, Manufactured for: Baxter…

    CGMP Deviations: Product was exposed to temperatures exceeding the labeled storage conditions during transportation were released by mistake.

    FDA · DrugBaxter Healthcare CorporationNationwide in the USAOfficial notice →
  9. 2023-10-04
  10. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Baxter Epiphany Cardio Server E3 ECG Management System Servers with software versions: a) v6.1.x, b) v6.2.x, c) v7.0.x.

    Epiphany, now a Baxter Healthcare company, is issuing an Urgent Medical Device Correction for the Cardio Server E3 ECG Management System software listed below due to a report of the caliper tool becoming…

    FDA · DeviceBaxter Healthcare CorporationWorldwide distribution - US Nationwide distribution including in the states of…Official notice →
  11. 2023-09-20
  12. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Welch Allyn Connex Spot Monitor

    Product is being recalled due to the improper placement of a copper tape on the back of the liquid-crystal display.

    FDA · DeviceBaxter Healthcare CorporationUS Nationwide distribution in the state of CA.Official notice →
All 123 Baxter Healthcare recalls →

Other drugs & medical recalls in the United States

  1. 2023-11-08
  2. US
    Drugs & medicalClass II

    Medtronic Neuromodulation recalls LEAD 3387S-40 STIMLOC DBS, Model Number 3387S-40. Electrical implantable device.

    There are a small number of units of the DBS Lead with StimLoc kit (3387S-40) due to a discrepancy in the Use By Date (UBD) printed on the outer packaged kit label versus the UBD printed on the individual product label.

    FDA · DeviceMedtronic NeuromodulationUS: MO, FL, TX, WAOfficial notice →
  3. US
    Drugs & medicalClass II

    Cepheid recalls Xpert Xpress Strep A, REF:Xprstrepa-CE-10, CE IVD,

    pipettes packaged in Strep A test may fail to dispense an adequate sample volume which could result in "invalid" results or no result test results.

    FDA · DeviceCepheidWorldwide distribution - US Nationwide and the countries of Costa Rica,…Official notice →
  4. US
    Drugs & medicalClass II

    Corin Ltd recalls Unity Total Knee System, Model Number 112.001.34. Primary and revision total knee arthroplasty

    Potential for Unity CR Inserts Right size 6 from batch 529803 to be incorrectly labelled as Unity CR Inserts Right size 7 from batch 532405 and visa versa.

    FDA · DeviceCorin LtdUS: MI, TX, ILOfficial notice →
  5. US
    Drugs & medicalClass II

    Cosmedent, Inc. recalls Cosmedent FlexiCup Composite Finishing & Polishing System, Six Pack, Autoclavable Aluminum Oxide Polishing Instrument, Medium (Blue), REF…

    Reports were received the elastomer cup material is crumbling.

    FDA · DeviceCosmedent, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  6. US
    Drugs & medicalClass II

    Maquet Cardiovascular, LLC recalls Vasoview 6 Pro (VV6) - Endoscopic Vessel Harvesting System, Model VH-2400 printed on outer label, inner label and IFU, outer box…

    Device packaged inside an incorrect outer box; displaying artwork & illustrations for a different device.

    FDA · DeviceMaquet Cardiovascular, LLCUS (AL, AR, CA, FL, GA, IN, KY, MI, NJ, OK, PA, TN, TX) and Canada.Official notice →
  7. US
    Drugs & medicalClass II

    Abbott recalls Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator

    Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.

    FDA · DeviceAbbottUS, Taiwan, and ColombiaOfficial notice →
See all drugs & medical recalls →