Drugs & medicalClass II

American Contract Systems Inc recalls Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ABDOMINAL HYSTERECTOMY - 205989…

Kits contain manufacturer-recalled perineal care (pericare) wipes within urinary catheter trays, which may be contaminated with a lubricant used during the manufacturing process. Use of a contaminated wipe may lead to…

Read the official notice →
Date
2026-10-07
Country
United States
Brand
American Contract Systems
Product
Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ABDOMINAL HYSTERECTOMY - 205989 Model…
Classification
Class II
Units
US Nationwide distribution in the states of Iowa, Minnesota, Wisconsin, and…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More American Contract Systems recalls

  1. 2026-09-02
  2. US
    Drugs & medicalClass II

    American Contract Systems Inc recalls American Contract System Convenience Kit, D & C PACK, Model/Catalog Number: LLPD90H

    The Convenience Kits contain Medline's Red Rubber Intermittent Catheters (8Fr-22 Fr) and Red Rubber Intermittent Catheter Trays. Medline Industries, LP has identified representative samples did not meet applicable…

    FDA · DeviceAmerican Contract Systems IncUS Nationwide distribution in the states of IA, OH.Official notice →
  3. US
    Drugs & medicalClass II

    American Contract Systems Inc recalls American Contract System Convenience Kit, CV B PACK, Model/Catalog Number: TNCB10AN

    The Convenience Kits contain Medline s Red Rubber Intermittent Catheters (8Fr-22 Fr) and Red Rubber Intermittent Catheter Trays. Medline Industries, LP has identified representative samples did not meet applicable…

    FDA · DeviceAmerican Contract Systems IncUS Nationwide distribution in the states of IA, OH.Official notice →
  4. 2026-08-26
  5. US
    Drugs & medicalClass II

    American Contract Systems Inc recalls American Contract Systems medical procedure kits and trays labeled as: 1) ANTERIOR CERVICAL PACK - 205988, Model Item Number…

    The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed¿ Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected…

    FDA · DeviceAmerican Contract Systems IncUS distribution to states of: MN, MO, IAOfficial notice →
  6. 2026-05-13
  7. US
    Drugs & medicalClass I

    American Contract Systems Inc recalls Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers…

    Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential…

    FDA · DeviceAmerican Contract Systems IncUS distribution to MN. No OUS distribution.Official notice →
  8. 2025-10-22
  9. US
    Drugs & medicalClass II

    American Contract Systems Inc. recalls Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991…

    Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to…

    FDA · DeviceAmerican Contract Systems Inc.US Nationwide distribution in the states of SD, IA, MN, WA, IL.Official notice →
  10. 2025-09-03
  11. US
    Drugs & medicalClass II

    American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ANTERIOR HIP PACK, MODEL/ITEM NUMBER AMAH40N; 2)TOTAL…

    During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS…

    FDA · DeviceAmerican Contract Systems IncUS distribution to states of: IL and LAOfficial notice →
All 22 American Contract Systems recalls →

Other drugs & medical recalls in the United States

  1. 2026-10-07
  2. US
    Drugs & medicalClass II

    Heraeus Medical GmbH (Dental Division) recalls PALACOS MV+G pro 40; PALACOS MV+G pro 80. Model/Catalog Number: 5081290; 5081291. Radiopaque, poly(methyl…

    Potential for monomer liquid leakage.

    FDA · DeviceHeraeus Medical GmbH (Dental Division)Worldwide distribution - US Nationwide and the countries of Aland Islands,…Official notice →
  3. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.

    There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.

    FDA · DeviceOlympus Corporation of the AmericasInternational distribution to the country of Japan.Official notice →
  4. US
    Drugs & medicalClass II

    Cardinal Health 200, LLC recalls Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT, GYN LAPAROSCOPY NEW HAN , Catalog Number: PB23GLNLAC; 2)…

    The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user…

    FDA · DeviceCardinal Health 200, LLCUS Nationwide distribution in the states of CA, IL, KY, MO, NC, NY, OH, TX.Official notice →
  5. US
    Drugs & medicalClass II

    American Nitrile Operations Llc recalls Brand Name: Performance Pro Product Name: Performance Pro Model/Catalog Number: 72-RM50601,72-RM50602, 72-RM50603,72-RM50604…

    Gloves may exhibit defects such as tears, missing fingers, and debris. Use of defective gloves may lead to healthcare personnel and patient exposure to infectious pathogens leading to infection or cross-contamination,…

    FDA · DeviceAmerican Nitrile Operations LlcUS Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Shanghai United Imaging Healthcare Co., Ltd. recalls Computed Tomography X-ray System, Model: uCT 530 with software

    Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images…

    FDA · DeviceShanghai United Imaging Healthcare Co., Ltd.US Nationwide distribution in the states of CA, PR, UT, WV, ND, NE, NC, TX,…Official notice →
  7. US
    Drugs & medicalClass II

    Volcano Corp recalls Philips IntraSight Plus REF: 797423 and 797417 100-120v-220-240v,50/60hZ, 1000 VA

    Software system may associate image data from one patient with another patient's record, potentially resulting in incorrect diagnosis or treatment.

    FDA · DeviceVolcano CorpWorldwide - US Nationwide distribution in the states of MA, MN, TX, AL, GA, CA…Official notice →
See all drugs & medical recalls →