Drugs & medicalClass II

Altimate Medical, Inc. recalls EasyStand Bantam, with the following sizes. a. Small; b. Extra Small. Used for assisting individual in standing.

Some EasyStand Bantam Small and Extra Small standers with the non-supine option may have been manufactured with a component that affects the operation of the leg assembly bracket to maintain the standers non-supine…

Read the official notice →
Date
2022-11-23
Country
United States
Brand
Altimate Medical
Product
EasyStand Bantam, with the following sizes. a. Small; b. Extra Small. Used for assisting individual in standing.
Classification
Class II
Units
Worldwide - US distribution in the states of CA, CT, FL, LA, MI, TN, TX and…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Altimate Medical recalls

  1. 2023-01-11
  2. US
    Drugs & medicalClass II

    Altimate Medical, Inc. recalls ActiveAid 922, Shower/Commode Chair

    Device manufactured with a component with affects the foldability of the frame, not allowing it to be secured by the frame strap in the folded position.

    FDA · DeviceAltimate Medical, Inc.US Nationwide distribution in the state of Texas.Official notice →
All 2 Altimate Medical recalls →

Other drugs & medical recalls in the United States

  1. 2022-11-23
  2. US
    Drugs & medicalClass II

    Novapproach Spine LLC recalls OneLIF Torque Limiting Adapter, AO to 1/4 SQ, 2.25Nm Limit, Catalog #015-010-0000, a component of NovApproach Spine OneLIF Interbody…

    The torque limiting adapter may not stay locked inside of the ratcheting handle once fully engaged.

    FDA · DeviceNovapproach Spine LLCUS Nationwide distribution in the states of CA and NY.Official notice →
  3. US
    Drugs & medicalClass II

    Tollos recalls Infection Control (IC) Lifting Strap used with the following Cirrus and Pinnacle Ceiling Lifts: Cirrus 450, Cirrus 600, Cirrus 700, Cirrus…

    Two customers reported four incidents in which the lifting strap broke while lifting a patient. If the IC strap fails during use, the patient may fall causing injury to them or the caregiver(s).

    FDA · DeviceTollosWorldwide - US Nationwide distribution and the countries of Canada and United…Official notice →
  4. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc. recalls Visualase Cooled Laser Applicator System (VCLAS) 9735560 .4MM CORE FIBER 10MM TIP product. Used to necrotize or coagulate soft tissue…

    Due to a defect in the outer pouch sterile seal

    FDA · DeviceMedtronic Navigation, Inc.U.S. Nationwide distribution including in the states AZ, CA, CO, DC, FL, IL,…Official notice →
  5. US
    Drugs & medicalClass II

    Texas Medical Technology Inc. recalls Disposable Surgical Level 3 Gown, Size/Product Code: Large/GWS-03-L-B, XLarge/GWS-03-XL-B, XXLarge/GWS-03-XXL-B

    Surgical gowns recalled due to gown fabric failing to pass the hydrostatic pressure test so the fabric does not provide the protection needed in a level 3 surgical gown, i.e., blood and other fluid may penetrate the…

    FDA · DeviceTexas Medical Technology Inc.US Nationwide distribution in the state of FL.Official notice →
  6. US
    Drugs & medicalClass II

    Steris Corporation recalls GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or…

    Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which…

    FDA · DeviceSteris CorporationUS Nationwide distribution in the states of Arizona, Illinois, Tennessee,…Official notice →
  7. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls Simplex HV With Gentamicin CE, 10 PACK-Bone cement with Gentomycin Catalog Number: 6193-1-010

    Simplex HV With Gentamicin CE, 10 PACKS (P/N 6193-1-010) were distributed to customers who had ordered Simplex HV With Gentamicin in US, 10 PACKS (P/N 6195-1-010)

    FDA · DeviceHowmedica Osteonics Corp.US Nationwide distribution.Official notice →