Aggredyne, Inc. recalls AggreGuide A-100 Instrument
The firm identified a software design issue where the improper shut down of the instrument resulted in speeds.csv file contents of the software version 5.1 being erased. Due to this issue, the target RPM for the assay…
Read the official notice →- Date
- 2022-09-07
- Country
- United States
- Brand
- Aggredyne, Inc.
- Product
- AggreGuide A-100 Instrument
- Classification
- Class II
- Units
- International distribution to the country of South Korea.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
Other drugs & medical recalls in the United States
- 2022-09-07
US
Medtronic Inc. recalls EverFlex Self-Expanding Peripheral Stent with Entrust Delivery System
Update to IFU provides a manual deployment workaround method to help mitigate potential harms related to partial stent deployment.
US
OXFORD IMMUNOTEC LTD recalls T-Cell Xtend REF TTK.610 US
Due to equipment failure, affected products were stored at temperatures outside the validated range of 2-8 degree C for a period longer than the validated storage time.
US
bioMerieux, Inc. recalls VITEK 2 / MYLA VITEK 2 Software versions 8.01, 8.02 (veterinary use only), 9.01, 9.02, 9.03 MYLA Software versions V4.8 and V4.9
There are 7 reported software anomalies that may affect use of the device.
FDA · DevicebioMerieux, Inc.International distribution to the countries of Estonia, Bahrain, Pakistan, UK,…Official notice →US
Aesculap Implant Systems LLC recalls Aesculap ENNOVATE SET SCREW STERILE-implants are used for dorsal monosegmental and multisegmental stabilization of the lumbar,…
Potential for the incorrect screw to be included in the package.
US
Philips North America Llc recalls DigitalDiagnost C50 1.1
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of the image on the first examination due to an issue in…
US
Medtronic Xomed, Inc. recalls NIM EMG Endotracheal Tube, Part Numbers: a) REF 8229308 b) REF 8229307 c) REF 8229306
The firm received reports of events related to airway obstruction while using NIM Standard Reinforced EMG Endotracheal Tubes & NIM CONTACT Reinforced EMG Endotracheal Tubes.
FDA · DeviceMedtronic Xomed, Inc.Worldwide distribution: US (nationwide) and OUS (foreign) countries of:…Official notice →