Drugs & medicalClass II

Aesculap Implant Systems LLC recalls Aesculap Caiman Articulating D5/360MM - An electrosurgical cutting and coagulation device for cutting/coagulation/ablation during…

Potential for the spring to become loose and fall out of the device, if the spring falls out during a surgical procedure, it may enter the patient's body.

Read the official notice →
Date
2023-05-17
Country
United States
Brand
Aesculap Implant Systems
Product
Aesculap Caiman Articulating D5/360MM - An electrosurgical cutting and coagulation device for cutting/coagulation/ablation during…
Classification
Class II
Units
US Nationwide distribution in the state of OH.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Aesculap Implant Systems recalls

  1. 2023-09-13
  2. US
    Drugs & medicalClass II

    Aesculap Implant Systems LLC recalls For intraventricular indications mainly used for operations within the brain ventricles. Product Code: FF399R

    Trocar manufactured with the shaft too long and does not meet manufacturing specifications. As a result, when used with the 30¿ neuroendoscope, a portion of the visual field is obscured.

    FDA · DeviceAesculap Implant Systems LLCUS Nationwide distribution in the states of FL, ID, IL, MD, NC, NY, OR, RI, TX.Official notice →
  3. 2023-07-26
  4. US
    Drugs & medicalClass II

    Aesculap Implant Systems LLC recalls Aesculap Surgical Needle, straight, 3 1/2", (90 mm), 13GA, width: 2,400 mm, non-sterile, reusable Article Number: MD610

    Incorrect product code marked on the product, etched with product code MD611, however; the correct product code is MD610.

    FDA · DeviceAesculap Implant Systems LLCUS Nationwide distribution in the states of DC, MO, PA, TX.Official notice →
  5. 2023-05-03
  6. US
    Drugs & medicalClass II

    Aesculap Implant Systems LLC recalls Caspar Cervical Retractor (CCR) Basket ME754 - Lid Only and ME764 Casper Cervical Retractor (CCR) Basket Full System - Lid Only. A…

    Mislabeled: Product is marked with GTIN # 04046955299592 however; the correct GTIN # is 04046955299607.

    FDA · DeviceAesculap Implant Systems LLCUS Nationwide distribution in the states of CA, OH, NE, NH, FL, PA, TX.Official notice →
  7. 2022-09-07
  8. US
    Drugs & medicalClass II

    Aesculap Implant Systems LLC recalls Aesculap ENNOVATE SET SCREW STERILE-implants are used for dorsal monosegmental and multisegmental stabilization of the lumbar,…

    Potential for the incorrect screw to be included in the package.

    FDA · DeviceAesculap Implant Systems LLCGA, TXOfficial notice →
  9. 2022-08-17
  10. US
    Drugs & medicalClass II

    Aesculap Implant Systems LLC recalls Aeos¿ Robotic Digital Microscope with software versions 2.7 or earlier Article Number: PV010

    Application error problems: Rebalance scope message: 1.Rebalancing process 2.Usage of Robot Recovery: 3. Autofocus not available:

    FDA · DeviceAesculap Implant Systems LLCUS Distribution to states of: AR, IN, NE and NCOfficial notice →
  11. 2022-08-10
  12. US
    Drugs & medicalClass II

    Aesculap Implant Systems LLC recalls PRESTIGE RETRACTION GRASPER, W/PORT DUAL ACTION, ROTATING 5mm DIAM., 36cm LENGTH. Product Code 8361-10

    Endoscopic graspers may become separated at the weld from handle to the shaft of the device.

    FDA · DeviceAesculap Implant Systems LLCUnited States Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO,…Official notice →
All 16 Aesculap Implant Systems recalls →

Other drugs & medical recalls in the United States

  1. 2023-05-17
  2. US
    Drugs & medicalClass II

    Bio-Rad Laboratories, Inc. recalls BioPlex 2200 REF 665-2050, APLS IgM Pack, APLS IgM Reagent Pack

    APLS IgM reagent kits were packaged with the incorrect conjugate, which could lead to an increase in false-positive and false-negative results.

    FDA · DeviceBio-Rad Laboratories, Inc.US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, FL,…Official notice →
  3. US
    Drugs & medicalClass II

    Ethicon, Inc. recalls PDS¿ Plus Antibacterial (polydioxanone) Suture - PDS PLUS CLR 27IN(70CM) USP3-0(M2) S/A PC-25 PRIME. Intended for use in general soft…

    Suture breakage during intra-operative use could result in poor performance of the impacted product because the intended benefit of tissue approximation and/or ligation may not be achieved

    FDA · DeviceEthicon, Inc.International distribution in the countries of United Kingdom, Ireland and…Official notice →
  4. US
    Drugs & medicalClass II

    New Wave Endo-Surgical, Corp. recalls M-Close Kit, REF 27-101; a disposable, ligature passing, suturing apparatus and needle guide for the abdominal wall

    The plastic housing on the device may fracture

    FDA · DeviceNew Wave Endo-Surgical, Corp.US Nationwide distribution in the states of FL, MA, and GA.Official notice →
  5. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls Bard Mission Disposable Core Biopsy Instrument Kit: REF: 2010MSK, 20G x 10CM; REF: 2016MSK, 20G x 16CM

    Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and…

    FDA · DeviceBard Peripheral Vascular IncUS Nationwide Distribution: PR, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA,…Official notice →
  6. US
    Drugs & medicalClass I

    Cordis US Corp recalls ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 501814REC

    There is a potential for separation of the ANGIOGUARD RX / XP delivery system and capture sheath which include but are not limited to situations of an intra-procedural delay, unplanned percutaneous or surgical…

    FDA · DeviceCordis US CorpWorldwide distribution.Official notice →
  7. US
    Drugs & medicalClass I

    Cook Incorporated recalls Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray, Reference Part Numbers C-PTISYJ-100-HC-G-EU-FLEX8.5 (G57725),…

    There were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in an unsecure…

    FDA · DeviceCook IncorporatedWorldwide distribution - US Nationwide.Official notice →