Drugs & medicalClass II

Accure Acne, Inc recalls Accure Laser System, Model: PFMS00004

Laser system's spot tracking system may become misaligned resulting in the possible risk of delivery of higher-than-expected laser energy to epidermal tissue, which may cause blistering.

Read the official notice →
Date
2024-01-31
Country
United States
Brand
Accure Acne, Inc
Product
Accure Laser System, Model: PFMS00004
Classification
Class II
Units
US distribution to states of: CO, CA, CT, NY, NJ, MN, LA, OH, MA, TX
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

Other drugs & medical recalls in the United States

  1. 2024-01-31
  2. US
    Drugs & medicalClass II

    SEDECAL SA recalls Sedecal X Optima URS digital mobile diagnostic x-ray system. A a stationary x-ray unit with a universal swivel arm. It allows one to take…

    Some parts have fallen off the equipment resulting in a safety risk to patients and/or users.

    FDA · DeviceSEDECAL SADistribution was made to AL, FL, and NC. There was no military/government…Official notice →
  3. US
    Drugs & medicalClass II

    Medtronic Neuromodulation recalls Model A710 Intellis Clinician Programmer Application, version 2.0.97

    Medtronic has discovered an instance in which the Intellis A710 Clinician Programmer Application (CP App) could no longer establish a session with the Implantable Neurostimulator (INS) which resulted in the inability…

    FDA · DeviceMedtronic NeuromodulationGermanyOfficial notice →
  4. US
    Drugs & medicalClass II

    Medtronic Inc. recalls Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator, product number DDPA2D1

    During internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the…

    FDA · DeviceMedtronic Inc.US Nationwide distribution in the state of Florida.Official notice →
  5. US
    Drugs & medicalClass II

    Turkuaz Saglik Hizmetleri Medikal Temizlik Kim. Ur. San.… recalls REF: 126590, MEDLINE, STERILE ULTRASOUND GEL, 0.70 FL oz. (20 mL), QTY: 50, STERILER, CE 2292

    Sterility of Ultrasound gel can not be guaranteed.

    FDA · DeviceTurkuaz Saglik Hizmetleri Medikal Temizlik Kim. Ur. San.…US distribution to: Illinois and OUS (foreign) distribution to: Hong KongOfficial notice →
  6. US
    Drugs & medicalClass II

    Tornier, Inc recalls Stryker Tornier Perform Reversed Glenoid Press-Fit Short Post Ti6AI4V, REF DWJ001, L: 7mm, Cementless Sterile.

    The incorrect device is contained in the labeled package.

    FDA · DeviceTornier, IncDistribution was made to AR, AZ, CA, CO, CT, NC, OH, TN, and VA. There was no…Official notice →
  7. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Anti-HBs Quantitative Reagent Pack, Catalog No. 6801925 VITROS Immunodiagnostic Products…

    Affected lots may experience increased calibration failures or an increase in falsely elevated results.

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Domestic distribution nationwide. OUS distribution to Bermuda, Australia,…Official notice →