Abiomed, Inc. recalls Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged,…
Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system…
Read the official notice →- Date
- 2026-09-30
- Country
- United States
- Brand
- Abiomed
- Product
- Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged,…
- Classification
- Class II
- Units
- Worldwide - US Nationwide and the countries of Austria, Belgium, Canada,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Abiomed recalls
- 2026-09-23
US
Abiomed, Inc. recalls Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged,…
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
FDA · DeviceAbiomed, Inc.Domestic: Nationwide Distribution. International: United Arab Emirates,…Official notice →- 2026-07-22
US
Abiomed, Inc. recalls ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr…
FDA · DeviceAbiomed, Inc.Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium…Official notice →- 2026-07-01
US
Abiomed, Inc. recalls Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits. Product Codes: 1000413,…
Potential for thrombus formation during prolonged use of the introducer.
FDA · DeviceAbiomed, Inc.Worldwide - US Nationwide distribution in the states of FL, IL, MN, NJ, NY, VA…Official notice →- 2026-06-24
US
Abiomed, Inc. recalls Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.
Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
FDA · DeviceAbiomed, Inc.US distribution to GA, IA, NE, PA, TX, WA, WI. International distribution to…Official notice →US
Abiomed, Inc. recalls Automated Impella Controller (AIC), for use with left heart support blood pump. Labeled as the following with corresponding Product Codes:…
Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.
FDA · DeviceAbiomed, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2026-06-03
US
Abiomed, Inc. recalls Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged,…
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper…
FDA · DeviceAbiomed, Inc.Worldwide distribution - US Nationwide and the countries of Canada, Denmark,…Official notice →
Other drugs & medical recalls in the United States
- 2026-09-30
US
DeRoyal Industries Inc recalls REF: 81-030415 Myocardial Probe Non-Pyrogenic , 400 Series
Myocardial temperature probes contain an incorrect needle. Specifically, an 18mm needle was included instead of a 15mm needle.
FDA · DeviceDeRoyal Industries IncUS Nationwide distribution in the states of LA,MO, MD, TN, UT.Official notice →US
Cardinal Health 200, LLC recalls Presource Convenience Kits: 1) KIT, PRE OP , Catalog Number: AMN548616A; 2) KIT, NICU ADMIT NEW HAN , Catalog Number: BA1DH0116B; 3)…
The kits contain applicators that may exhibit an open or incomplete seal on the packaging of the applicator. Use of a potentially contaminated applicator could result in no patient harm or may lead to localized skin or…
FDA · DeviceCardinal Health 200, LLCUS Nationwide distribution in the states of FL, KY, MO, NC, OH, TX.Official notice →US
Criticare Technologies, Inc. recalls nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US,…
Complaint trend where "Checkline Error" alarm presents at device setup, rendering the CO2 monitoring function inoperable. May result in procedure delay.
FDA · DeviceCriticare Technologies, Inc.US Nationwide distribution including in the states and territory of AL, AZ,…Official notice →US
Becton, Dickinson and Company recalls ForeSight Elite Tissue Oximetry Cable REF HEMFSM10 UDI-DI code: 00690103210927 The non-invasive ForeSight oximeter cable is…
Due to incorrect switching of Channel 1 and 2 on the module and flag labels on the tissue oximetry cables.
FDA · DeviceBecton, Dickinson and CompanyWorldwide - U.S Nationwide distribution including in the states of AK, AL, AR,…Official notice →US
Philips Ultrasound, LLC recalls Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234
Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.
FDA · DevicePhilips Ultrasound, LLCWorldwide - US Nationwide distribution in the states of AL, CA, CT, FL, GA,…Official notice →US
CooperVision, Inc. recalls Brand Name: Pearle Vision 1 day Product Name: (somofilcon A) Contact Lens Model/Catalog Number: Pearle Vision 1 day
A small number of lenses within the affected lot are of the incorrect power.
FDA · DeviceCooperVision, Inc.US Nationwide distribution in the states of MN, OH, WI.Official notice →