Drugs & medicalClass II

Abbott recalls Ultimum HEMOSTASIS INTRODUCER 5f 1.67 MM

As a result of a manufacturing error, one (1) lot of Ultimum Hemostasis Introducer devices was manufactured with an incorrect hub and overall length. The devices manufactured were 12 cm in length instead of the…

Read the official notice →
Date
2022-01-05
Country
United States
Brand
Abbott
Product
Ultimum HEMOSTASIS INTRODUCER 5f 1.67 MM
Classification
Class II
Units
Worldwide distribution - US Nationwide distribution in the state of Missouri…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abbott recalls

  1. 2022-06-19
  2. CA
  3. 2022-06-08
  4. US
    Drugs & medicalClass II

    Abbott recalls FIRMap" Catheter, 60mm Basket

    Incorrect product labeling.

    FDA · DeviceAbbottUS Nationwide Distribution to states of: MO, CA, OH, WA and OUS…Official notice →
  5. 2022-04-27
  6. US
    Drugs & medicalClass II

    Abbott recalls Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve DD, Model A-TCSE-DD, Sterile.

    Devices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device.

    FDA · DeviceAbbottWorldwide distribution - US Nationwide - there was government distribution but…Official notice →
  7. 2022-04-13
  8. US
    Drugs & medicalClass II

    Abbott recalls Amplatzer Steerable Delivery Sheath

    Instructions for use were updated.

    FDA · DeviceAbbottOUS distribution onlyOfficial notice →
  9. 2022-02-20
  10. UK
  11. 2022-02-17
  12. CA
All 15 Abbott recalls →

Other drugs & medical recalls in the United States

  1. 2022-01-05
  2. US
    Drugs & medicalClass II

    Forbes Rehab Services Inc recalls Affected components are limited to the Monoprice charging cable (Product # 4867) and 24W iClever Boost Cube charging brick components…

    Warning Added: Using damaged cables or chargers, or charging when moisture is present, can cause fire, electric shock, injury, or damage to Speech Generating Device or other property

    FDA · DeviceForbes Rehab Services IncUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Biomerieux Inc recalls RAPID 20 E, REF 20701, System for the identification of Enterobacteriaceae in 4 hours, 25 strips/package, IVD.

    There is an error in the new version of the package insert. The Reading Table lists the incorrect color for a negative result for the CIT test.

    FDA · DeviceBiomerieux IncUS Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA,…Official notice →
  4. US
    Drugs & medicalClass II

    Migo Trading LLC recalls FLOWFLEX SARS-COV-2 Antigen Rapid Test (Self-Testing), 1 Test

    Devices are determined to be inconsistent with the firm's Emergency Use Authorization (EUA). This product has not been authorized for emergency use and the test cannot be determined to be adequately validated for the…

    FDA · DeviceMigo Trading LLCNYOfficial notice →
  5. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls da Vinci SP Surgical systems with the following Model Name/ Model Number / UDI: ASSY,PSS,SP1098,P4/ 380601-44/ 00886874114605…

    Issue was identified during internal engineering evaluation. Use of the system with the affected software version may experience either an inability to deliver energy or an inadvertent delivery of energy. Use may…

    FDA · DeviceIntuitive Surgical, Inc.US Nationwide distribution in the states of Arizona, California, Connecticut,…Official notice →
  6. US
    Drugs & medicalClass II

    Medtronic Neuromodulation recalls Model B35200 - Percept PC BrainSense Implantable Neurostimulator

    During cardioversion events, the Model B35200 Percept PC Implantable Neurostimulator could become unresponsive and non-functional. Cardioversion is a procedure performed by a cardiologist to correct an abnormal…

    FDA · DeviceMedtronic NeuromodulationWorldwide distribution - US Nationwide and Puerto Rico. The countries of…Official notice →
  7. US
    Drugs & medicalClass I

    Valisa MFG LLC recalls kleantouch HAND SANITIZER (ethyl alcohol 70%), 8.0 FL OZ (236 ML), pump bottles, Distributed by: Valisa MFG, LLC, Farmingdale, NY, 11735,…

    Marketed Without an Approved NDA/ANDA: Product labeled to contain methanol making it an unapproved new drug and contains methanol

    FDA · DrugValisa MFG LLCNationwide in the USAOfficial notice →
See all drugs & medical recalls →