Drugs & medicalClass II

Siemens Healthcare Diagnostics Inc recalls Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 500e Blood Gas…

Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software…

Read the official notice →
Date
2024-07-03
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 500e Blood Gas System (China)…
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of FL, GA, LA, MI, NC and…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2024-09-18
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated…

    Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distributionOfficial notice →
  3. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista¿ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;

    During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Domestic: Nationwide Distribution. International: Austria, Belgium, Canada,…Official notice →
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP); Siemens Material Number (SMN): 11097543

    The potential for biased quality control (QC) and patient results when using Atellica CH UCFP lot 130414 and ADVIA Chemistry UCFP lot 140 reagents.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of CA, FL, GA, IA, IL,…Official notice →
  5. 2024-05-29
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 Anti-TG Ab Catalog # L2KTG2 (200 Tests), SMN 10381659, System, Test, Thyroid Autoantibody.

    The potential for falsely elevated patient sample results when using impacted lots (782, 783, 785, 786, 787, 788, 789, 790, 791, 792 and 793), of IMMULITE 2000 Anti-TG Ab.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution including in the states of AL, CA, DC,…Official notice →
  7. 2024-04-24
  8. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparin and sodium…

    Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positive bias ranging from 2-16% -impacts…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  9. 2024-03-27
  10. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls DCA Systems HbA1c Reagent Kit

    The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the country of Taiwan.Official notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2024-07-03
  2. US
    Drugs & medicalClass II

    Merit Medical Systems, Inc. recalls Prelude Roadster Guide Sheath REF: -PG4F45S -PG4F90S -PG5F45MP -PG5F45R -PG5F45S -PG5F45S018 -PG5F90MP -PG5F90S -PG5F90S018 -PG6F45HS…

    Due to older version of Instruction for Use (IFU) instructing to ensure that the value and the sheath are tightly connected before use. This may cause users to over-tightening the luer which may result in the swivel…

    FDA · DeviceMerit Medical Systems, Inc.Worldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →
  3. US
    Drugs & medicalClass II

    NovaSignal Corp. recalls NovaGuide 2 Traditional Ultrasound , REF: NSC-TCDNG2

    Due to inadvertently disconnecting grounding cable of ultrasound system could result in electrical shock, electrical interference and/or equipment malfunction.

    FDA · DeviceNovaSignal Corp.Worldwide - US Nationwide distribution in the states of MA, TX, NY, MN, AZ,…Official notice →
  4. US
    Drugs & medicalClass II

    Thoratec Switzerland GMBH recalls 2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L201-90401, L201-90411,…

    Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down…

    FDA · DeviceThoratec Switzerland GMBHWorldwide - US Nationwide distribution included in the states of CA, PA, FL,…Official notice →
  5. US
    Drugs & medicalClass II

    Mermaid Medical A/S recalls The D*Clot¿ HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system…

    Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.

    FDA · DeviceMermaid Medical A/SWorldwide - US Nationwide distribution in the states of OH, MO, CO, NC, TX,…Official notice →
  6. US
    Drugs & medicalClass II

    Medtronic Inc. recalls EVERA MRI S VR SureScan, Model Number DVMC3D4, Implantable Cardioverter Defibrillator

    Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.

    FDA · DeviceMedtronic Inc.Worldwide - US Nationwide distribution in the states of CA, CT, FL, IN, MA,…Official notice →
  7. US
    Drugs & medicalClass II

    Draeger Medical Systems, Inc. recalls Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23

    Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and…

    FDA · DeviceDraeger Medical Systems, Inc.US Nationwide distribution. International distribution to Afghanistan,…Official notice →
See all drugs & medical recalls →