Mermaid Medical A/S recalls The D*Clot¿ HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system…
Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.
Read the official notice →- Date
- 2024-07-03
- Country
- United States
- Brand
- Mermaid Medical A/S
- Product
- The D*Clot¿ HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system…
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of OH, MO, CO, NC, TX,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
Other drugs & medical recalls in the United States
- 2024-07-03
US
NovaSignal Corp. recalls NovaGuide 2 Traditional Ultrasound , REF: NSC-TCDNG2
Due to inadvertently disconnecting grounding cable of ultrasound system could result in electrical shock, electrical interference and/or equipment malfunction.
FDA · DeviceNovaSignal Corp.Worldwide - US Nationwide distribution in the states of MA, TX, NY, MN, AZ,…Official notice →US
Thoratec Switzerland GMBH recalls 2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L201-90401, L201-90411,…
Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down…
FDA · DeviceThoratec Switzerland GMBHWorldwide - US Nationwide distribution included in the states of CA, PA, FL,…Official notice →US
Siemens Healthcare Diagnostics Inc recalls Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 500e Blood Gas…
Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software…
FDA · DeviceSiemens Healthcare Diagnostics IncWorldwide - US Nationwide distribution in the states of FL, GA, LA, MI, NC and…Official notice →US
Medtronic Inc. recalls EVERA MRI S VR SureScan, Model Number DVMC3D4, Implantable Cardioverter Defibrillator
Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.
FDA · DeviceMedtronic Inc.Worldwide - US Nationwide distribution in the states of CA, CT, FL, IN, MA,…Official notice →US
Draeger Medical Systems, Inc. recalls Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23
Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and…
FDA · DeviceDraeger Medical Systems, Inc.US Nationwide distribution. International distribution to Afghanistan,…Official notice →US
Siemens Medical Solutions USA, Inc recalls AXIOM Vertix MD - The Vertix MD Trauma has been specially designed for examining emergency and accident patients as well as…
Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire