Drugs & medicalClass II

Qiagen Sciences LLC recalls QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223

The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both…

Read the official notice →
Date
2022-01-12
Country
United States
Brand
Qiagen Sciences
Product
QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223
Classification
Class II
Units
Domestic distribution to AL, DC, KY, MD, ME, NJ, OK< PA, and TX. No foreign…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Qiagen Sciences recalls

  1. 2024-07-24
  2. US
    Drugs & medicalClass II

    Qiagen Sciences LLC recalls QIAcube Connect MDx - IVD Instrument designed to perform automated isolation and purification of nucleic acids in molecular diagnostic…

    Heater shaker module does not perform heating if the temperature is set below 40¿C for static incubation steps when running IVD protocol QIAamp¿ DSP DNA Mini Kit (cat. no. 61304), Protocol: Isolation of genomic DNA…

    FDA · DeviceQiagen Sciences LLCNationwide. Foreign: AD, AT, AU, BE, CA, CH, CN, CZ, DE, DK, ES, FI, FR, GB,…Official notice →
  3. 2024-04-17
  4. US
    Drugs & medicalClass II

    Qiagen Sciences LLC recalls EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or molecular…

    Insufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER UNABLE TO REACH TEMPERATURE; the instrument may stop, and the sample being processed will be lost

    FDA · DeviceQiagen Sciences LLCNationwide Foreign: AT, AU, BE, CH, CR, CZ, DE, DK, ES, FR, GB, HU, IT, JO,…Official notice →
  5. 2023-06-14
  6. US
    Drugs & medicalClass II

    Qiagen Sciences LLC recalls QIAstat-Dx Respiratory SARS-CoV-2 Panel multiplexed nucleic acid real-time PCR test, Reference Number 691223

    Faulty cartridges in the lot could result in false test results.

    FDA · DeviceQiagen Sciences LLCUS Nationwide distribution in the states of CA, FL, IL, MD, MS, NY, OK, PA, TN.Official notice →
  7. 2022-09-21
  8. US
    Drugs & medicalClass II

    Qiagen Sciences LLC recalls QIAcube Connect MDx, Model No. 9003070

    During the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is instead the exact volume that should be loaded. Misunderstanding the…

    FDA · DeviceQiagen Sciences LLCDomestic distribution to AZ, CA, CO, DE, DC, FL, GA, IA, KS, LA, MA, MD, MN,…Official notice →
  9. 2022-09-07
  10. US
    Drugs & medicalClass II

    Qiagen Sciences LLC recalls FGFR RGQ RT-PCR Kit (RUO, not IVD) REF 8747010

    Increased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive result could lead to the risk of the…

    FDA · DeviceQiagen Sciences LLCIN, NC, MN, TX Foreign: Belgium, China, Germany, Jordan, Italy, Netherlands,…Official notice →
  11. 2022-04-27
  12. US
    Drugs & medicalClass II

    Qiagen Sciences LLC recalls therascreen KRAS RGQ PCR Kit (24) Model Number(s): REF 870021 (US IVD)

    False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutation target of the kit.

    FDA · DeviceQiagen Sciences LLCAZ CA CT FL GA IL IN MA MN NC NY OK OR PA TN TX WVOfficial notice →
All 8 Qiagen Sciences recalls →

Other drugs & medical recalls in the United States

  1. 2022-01-12
  2. US
    Drugs & medicalClass II

    Uromedica Inc. recalls UROMEDICA Implantation Instrument Set (Sheath and Sharp Trocar), REF 750041, nonsterile.

    The U-channel sheaths may cause damage to ProACT implant devices when the devices are passed through the sheath during a ProACT implantation procedure.

    FDA · DeviceUromedica Inc.US Nationwide distribution in the states of AZ, CA, CO, KS, MD, TX, and WI.Official notice →
  3. US
    Drugs & medicalClass II

    DiaSorin Molecular LLC recalls Simplexa Direct Amplification Disc Kit, Product Kit Code MOL1455

    A subset of affected discs are at risk of leakage when it is used more than one time with Simplexa COVID-19 Direct and/or Simplexa Flu A/B & RSV Direct Gen II. Leakage from a well that had a positive sample may cause…

    FDA · DeviceDiaSorin Molecular LLCUS: AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI,…Official notice →
  4. US
    Drugs & medicalClass II

    Nikkiso Ltd - Shizuoka Plant recalls DBB-06 Hemodialysis Delivery System

    Device lacks premarket approval.

    FDA · DeviceNikkiso Ltd - Shizuoka PlantDistributed nationwide to NY, MO, WA, CA, and TN.Official notice →
  5. US
    Drugs & medicalClass III

    Tapemark Company recalls iontoPATCH On-the-Go Patch Therapy, Transdermal Iontophoretic Drug Delivery System, Extra Strength, 8-HOUR Average Wear Time, 6/1-patch…

    The dosage information on the pouch is incorrect; the primary carton label is correct.

    FDA · DeviceTapemark CompanyUS Nationwide distribution in the states of CA, FL, IL, MO, NY, OH, and TN.Official notice →
  6. US
    Drugs & medicalClass II

    Medshape, INC. recalls DYNANAIL MINI, Endcap +6mm, REF 2600-05-006, Rx Only, Sterile

    The firm identified a nonconformance in Lot 04882 involving the endcap being manufactured to a longer length than indicated in product specifications, introducing the risk of the +6 mm endcap not seating flush with the…

    FDA · DeviceMedshape, INC.Domestic: CA, FL, GA, IL, IN, MA, MD, TX, and WA.Official notice →
  7. US
    Drugs & medicalClass II

    Biomeme, Inc. recalls Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555;…

    The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once…

    FDA · DeviceBiomeme, Inc.Domestic distribution to AZ, CA, FL, HI, IN, KY, MD, MO, NC, NJ, NY, OH, OK,…Official notice →
See all drugs & medical recalls →