Qiagen Sciences LLC recalls EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or molecular…
Insufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER UNABLE TO REACH TEMPERATURE; the instrument may stop, and the sample being processed will be lost
Read the official notice →- Date
- 2024-04-17
- Country
- United States
- Brand
- Qiagen Sciences
- Product
- EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or molecular…
- Classification
- Class II
- Units
- Nationwide Foreign: AT, AU, BE, CH, CR, CZ, DE, DK, ES, FR, GB, HU, IT, JO,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Qiagen Sciences recalls
- 2025-04-09
US
Qiagen Sciences LLC recalls Brand Name: QIAstat-Dx Product Name: QIAstat-Dx Respiratory Panel Plus Reference Number (REF): 691224 Software Version: n/a Component: n/a
Identified faulty cartridges. If such a faulty cartridge is used for sample testing, false test results could occur.
FDA · DeviceQiagen Sciences LLCUS Distribution: AL, CA, DC, FL, IN, KS, KY, MA, ME, MN, NY, and TX.Official notice →- 2024-07-24
US
Qiagen Sciences LLC recalls QIAcube Connect MDx - IVD Instrument designed to perform automated isolation and purification of nucleic acids in molecular diagnostic…
Heater shaker module does not perform heating if the temperature is set below 40¿C for static incubation steps when running IVD protocol QIAamp¿ DSP DNA Mini Kit (cat. no. 61304), Protocol: Isolation of genomic DNA…
FDA · DeviceQiagen Sciences LLCNationwide. Foreign: AD, AT, AU, BE, CA, CH, CN, CZ, DE, DK, ES, FI, FR, GB,…Official notice →- 2023-06-14
US
Qiagen Sciences LLC recalls QIAstat-Dx Respiratory SARS-CoV-2 Panel multiplexed nucleic acid real-time PCR test, Reference Number 691223
Faulty cartridges in the lot could result in false test results.
FDA · DeviceQiagen Sciences LLCUS Nationwide distribution in the states of CA, FL, IL, MD, MS, NY, OK, PA, TN.Official notice →- 2022-09-21
US
Qiagen Sciences LLC recalls QIAcube Connect MDx, Model No. 9003070
During the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is instead the exact volume that should be loaded. Misunderstanding the…
FDA · DeviceQiagen Sciences LLCDomestic distribution to AZ, CA, CO, DE, DC, FL, GA, IA, KS, LA, MA, MD, MN,…Official notice →- 2022-09-07
US
Qiagen Sciences LLC recalls FGFR RGQ RT-PCR Kit (RUO, not IVD) REF 8747010
Increased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive result could lead to the risk of the…
FDA · DeviceQiagen Sciences LLCIN, NC, MN, TX Foreign: Belgium, China, Germany, Jordan, Italy, Netherlands,…Official notice →- 2022-04-27
US
Qiagen Sciences LLC recalls therascreen KRAS RGQ PCR Kit (24) Model Number(s): REF 870021 (US IVD)
False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutation target of the kit.
Other drugs & medical recalls in the United States
- 2024-04-17
US
Intuitive Surgical, Inc. recalls Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098
Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries,…
FDA · DeviceIntuitive Surgical, Inc.US: VA, CA, NH, FL, NY, CO, AL, KY, MD, NC, NJ, OH, SC, SD, TN, WA, WI, WV,…Official notice →US
Boston Scientific Corporation recalls EndoVive 20 Fr Safety PEG Push w/ENFit Kit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and…
Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a…
FDA · DeviceBoston Scientific CorporationWorldwide distribution - US Nationwide and the country of Japan.Official notice →US
Baxter Healthcare Corporation recalls Hillrom Welch Allyn Connex ProBP 3400 Digital Blood Pressure Device, marketed under the following models: a) PROBP 3400, SUREBP,…
Baxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and international electrical standards.
US
Philips North America recalls PHILIPS Incisive Computed tomography X-ray system Incisive CT Software Upgrade from V4.5.1.41921 to V4.5.1.41989-Intended to produce…
Patient tabletop moved out to the home position during scan initialization, may cause operator/bystander staying by the rear of the table to collide with the moved tabletop and be injured
FDA · DevicePhilips North AmericaNationwide including Puerto Rico Foreign: Albania Argentina Australia Austria…Official notice →US
Stryker Communications recalls Chromophare Surgical Light System , REF CH00000001
Due to insufficient mounting force the stability of surgical light system cannot be guaranteed.
US
Synovo Production recalls Femoral Resurfacing Cup - (xx)MM, Ti-6AI-4V ELI, TiN CERAMIC W/ PLASMA SPRAY, STERILEEO, Reference Numbers: 01-03-0042 01-03-0044…