Physio-Control recalls

11 official recalls and safety alerts involving Physio-Control since 2022-03, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Physio-Control or the relevant agency.

  1. 2026-05-13
  2. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957,…

    Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's…

    FDA · DevicePhysio-Control, Inc.US Nationwide distribution in the states of MI, WA, KY, IN, FL, NC, TX, LA,…Official notice →
  3. 2026-03-18
  4. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog…

    Due to pediatric defibrillator electrode delamination

    FDA · DevicePhysio-Control, Inc.Worldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →
  5. 2025-09-10
  6. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls LIFEPAK CR2 Part / Catalog Numbers / UDI-DI code: CR2-2-001451 99512-001434 883873785911 LIFEPAK CR2: The LIFEPAK CR2 defibrillator is…

    Due to required inspections not being performed on products/units that have gone through servicing.

    FDA · DevicePhysio-Control, Inc.U.S. Nationwide distribution in the states of CO and MT.Official notice →
  7. 2025-03-05
  8. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.

    Incorrect keypad was incorrectly installed onto a defibrillator/monitor.

    FDA · DevicePhysio-Control, Inc.U.S.: WV O.U.S.: N/AOfficial notice →
  9. 2025-02-26
  10. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo¿ RD rainbow Adt 8¿ SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo¿…

    Due to an error message that prevents users from utilizing carboxyhemoglobin saturation and methemoglobin saturation sensors on their monitor/defibrillator system.

    FDA · DevicePhysio-Control, Inc.U.S. Nationwide distribution including in the states of CA, CO, CT, FL, GA,…Official notice →
  11. 2025-02-12
  12. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls LIFEPAK 35 ECG cable REF 11111-000041

    Product shipped with incorrect IFU. IFU does not include proper cleaning/disinfection instructions

    FDA · DevicePhysio-Control, Inc.U.S. Nationwide distribution in the states of CA, FL, GA, IA, IN, KY, MD, MI,…Official notice →
  13. 2024-11-06
  14. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o…

    Due to an out of tolerance tool being used on monitor/defibrillator systems.

    FDA · DevicePhysio-Control, Inc.U.S.: CT, GA, MA, NY, OH, WA, and WV. O.U.S.: N/AOfficial notice →
  15. 2024-10-30
  16. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls STERILIZABLE INTERNAL DEFIBRILLATION PADDLES REF LP20 STERILIZABLE INTERNAL DEFIBRILLATION PADDLES REF LP15 Defibrillation is indicated…

    Due to customer complaints and out of box failures, the firm is conducting a voluntary recall as a result of product failures (cracks and over-molding separation of the sterilizable internal defibrillation paddles.

    FDA · DevicePhysio-Control, Inc.Worldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →
  17. 2023-12-27
  18. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls AC Power Adapter for use with the LIFEPAK 15 Monitor/Defibrillator, REF: 11140-000098

    AC Power Adapter (ACPA) may not charge the monitor/defibrillator batteries and may not allow the monitor/defibrillator to power on while plugged into the ACPA.

    FDA · DevicePhysio-Control, Inc.US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO,…Official notice →
  19. 2023-11-22
  20. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls LIFEPAK 15, REF: 99577-000046, Part Number V15-2-000051

    Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.

    FDA · DevicePhysio-Control, Inc.US Nationwide distribution in the states of AZ, CA, NC, NJ, NY.Official notice →
  21. 2022-03-16
  22. US
    Drugs & medicalClass III

    Physio-Control, Inc. recalls HeartSine Samaritan PAD, SAM 350P Semi-Automatic Defibrillator, used with Pad-Pak

    Semi-automatic defibrillators shipped out by Stryker without battery and electrode packs.

    FDA · DevicePhysio-Control, Inc.U.S.: AKOfficial notice →