Drugs & medicalClass II

Terumo Cardiovascular Systems Corporation recalls Cardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974 65246 66312…

Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially requiring surgical or…

Read the official notice →
Date
2024-04-10
Country
United States
Brand
Terumo Cardiovascular Systems
Product
Cardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974 65246 66312 66342 66388 66650…
Classification
Class II
Units
worldwide distribution - US Nationwide and the countries of Canada, Chile,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Terumo Cardiovascular Systems recalls

  1. 2025-07-02
  2. US
    Drugs & medicalClass II

    Terumo Cardiovascular Systems Corporation recalls CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial…

    Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in…

    FDA · DeviceTerumo Cardiovascular Systems CorporationUS Nationwide distribution in the states of California, Delaware, Hawaii,…Official notice →
  3. 2025-04-09
  4. US
    Drugs & medicalClass II

    Terumo Cardiovascular Systems Corporation recalls Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754

    During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be…

    FDA · DeviceTerumo Cardiovascular Systems CorporationUS: IL, TX, NY, MI, FL, NV, and Australia, New ZealandOfficial notice →
  5. 2023-04-12
  6. US
    Drugs & medicalClass II

    Terumo Cardiovascular Systems Corporation recalls CAPIOX NX19 Hollow Fiber Oxygenator, Catalog Number 3CX*NX19RE

    Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly…

    FDA · DeviceTerumo Cardiovascular Systems CorporationUS Nationwide. Belgium, Canada, Costa Rica, Thailand.Official notice →
  7. 2022-04-27
  8. US
    Drugs & medicalClass II

    Terumo Cardiovascular Systems Corp recalls Cardiovascular Procedure Kit catalog # 76645 & 73806

    Potential for Cardiovascular Procedure Kit packaging damage that occurred during shipping.

    FDA · DeviceTerumo Cardiovascular Systems CorpUS Nationwide distribution in the states of CA and IN.Official notice →
All 5 Terumo Cardiovascular Systems recalls →

Other drugs & medical recalls in the United States

  1. 2024-04-10
  2. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Centurion manual surgical kits labeled as: a) STERILE 4-1/2" S/S SCISSOR (SS5031), Product Code 65145; b) STERILE S/B…

    Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Panama, Canada.Official notice →
  3. US
    Drugs & medicalClass II

    Life Technologies Corporation recalls Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems-IVD intended for targeted…

    Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems exploitation of the vulnerability by a threat actor may allow them to alter settings, configurations, software, or data on…

    FDA · DeviceLife Technologies CorporationUS Nationwide distribution in the states of CA, IN, NC OR, TN.Official notice →
  4. US
    Drugs & medicalClass II

    Exactech, Inc. recalls Exactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 12o, RIGHT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item…

    The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional…

    FDA · DeviceExactech, Inc.Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China,…Official notice →
  5. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Hill-Rom PRO+ 36" MRS Surface, REF P7924A03, The pro+ mattress is intended for patient support and for the prevention and/or…

    Inconsistencies were identified with service records associated with corrections performed for a previous Medical Device Correction issued by Baxter (ref number: FA-2022-026). Your devices have been identified as…

    FDA · DeviceBaxter Healthcare CorporationUS Nationwide distribution in the state of Colorado.Official notice →
  6. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray)…

    Esculin, a substrate used in other panels and normally in A16 well, was found in A17 well. A17 in yeast panels is supposed to be an empty well and is not a part of the normal ID evaluation. However, A17 is checked for…

    FDA · DeviceBecton Dickinson & Co.Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN,…Official notice →
  7. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Phoenix NMIC-306 Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray)…

    Products were found to have confirmed drug cross contamination prior to release may lead to incorrect, inadequate, or absent antibiotic treatment or delay in results

    FDA · DeviceBecton Dickinson & Co.Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN,…Official notice →
See all drugs & medical recalls →