Drugs & medicalClass II

Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;

Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low…

Read the official notice →
Date
2026-05-06
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
Classification
Class II
Units
Worldwide distribution - US Nationwide and the countries of Italy, Spain,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2026-08-12
  2. US
    Drugs & medicalClass III

    Siemens Healthcare Diagnostics, Inc. recalls Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL

    Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of CA and VA. The country…Official notice →
  3. 2026-07-15
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.

    The potential for intermittent well-to-well bias on the A1c_E assay on the Atellica CH and Atellica CI analyzer which could cause falsely depressed or falsely elevated HbA1c patient results.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  5. 2026-06-17
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay

    A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  7. 2026-05-06
  8. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.

    Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  9. 2026-04-08
  10. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;

    A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Domestic: Nationwide Distribution International: Argentina, Australia,…Official notice →
  11. 2026-03-11
  12. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urine Albumin (UAlb). Material Number: 11537225

    Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2026-05-06
  2. US
    Drugs & medicalClass II

    Medline Industries, LP recalls Medline medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: 1. BLOOD CULTURE KIT, Medline Kit SKU DYNDH1520;…

    Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be…

    FDA · DeviceMedline Industries, LPUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Medline Industries, LP recalls Medline Neuro Sponges, various dimensions, labeled as: 1. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.25" X 0.25", XR, 10/PK,…

    During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

    FDA · DeviceMedline Industries, LPWorldwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Elekta, Inc. recalls Brand Name: Leksell Vantage Arc System REF: 1053958

    Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.

    FDA · DeviceElekta, Inc.US: AR AZ CA CO DC FL GA ID IL KS MA MD ME MO MT NC NE NH NJ NM NV NY OH OK PA…Official notice →
  5. US
    Drugs & medicalClass II

    Merit Medical Systems, Inc. recalls Philips Bridge Prep Kit REF: K12-09098C Sterile EO, Rx ONLY

    catheter may experience resistance when being advanced over the guidewire

    FDA · DeviceMerit Medical Systems, Inc.US Nationwide distribution in the state of CO.Official notice →
  6. US
    Drugs & medicalClass II

    Abbott Point Of Care Inc. recalls i-STAT G3+ cartridge; List Number: 03P78-26;

    Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than…

    FDA · DeviceAbbott Point Of Care Inc.Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA,…Official notice →
  7. US
    Drugs & medicalClass III

    Heraeus Medical GmbH (Dental Division) recalls Heraeus, PALAMIX uno. Material Number: 66057893.

    Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce…

    FDA · DeviceHeraeus Medical GmbH (Dental Division)Worldwide distribution - US Nationwide and the countries of Aland Islands,…Official notice →
See all drugs & medical recalls →