Randox Laboratories Ltd. recalls RX Series CYSC- Cystatin C Reagent -Intended for in vitro diagnostic (IVD) use as Randox Cystatin C assay on automated analysers.…
Cystatin C Reagent marketed without a 510 (k)
Read the official notice →- Date
- 2024-07-31
- Country
- United States
- Brand
- Randox Laboratories
- Product
- RX Series CYSC- Cystatin C Reagent -Intended for in vitro diagnostic (IVD) use as Randox Cystatin C assay on automated analysers.…
- Classification
- Class II
- Units
- US Nationwide distribution in the states of AL, CA, GA, NY.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Randox Laboratories recalls
- 2025-01-29
US
Randox Laboratories, Limited recalls Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catalog Number: LP2757
Lipoprotein (a) on Roche Cobas c501 with nmol/l claims does not have a 510(k) or PMA.
- 2024-05-15
US
Randox Laboratories Ltd. recalls Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and metabolites of…
Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This…
US
Randox Laboratories Ltd. recalls Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used…
Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.
- 2024-04-24
US
Randox Laboratories Ltd. recalls Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample…
FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of GA, MA, WV including PR.Official notice →- 2024-04-03
US
Randox Laboratories Ltd. recalls Cystatin C Calibrator - In vitro diagnostic (IVD) use in the calibration of the Randox Cystatin C assay on automated analysers Catalog…
Cystatin C Calibrator Series CVS 2699, is running with a negative bias compared to other methods
FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of CA, IN.Official notice →- 2024-03-27
US
Randox Laboratories Ltd. recalls Liquid Protein Calibrators. C3 and Haptoglobin liquid protein calibrators used with Randox IgA, IgG, and IgM assays that require sample…
The firm has realigned C3 and Haptoglobin in Liquid Protein Calibrators, IT2691, to reference material ERM-DA470k/IFCC.
Other drugs & medical recalls in the United States
- 2024-07-31
US
Beckman Coulter Inc. recalls Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2
LIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to report as normal
FDA · DeviceBeckman Coulter Inc.Worldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →US
Medtronic Neuromodulation recalls A610 Clinician Programmer (CP) Software Application (app) with the following versions: Version 2.0.x (2.0.4584, 2.0.4594, 2.0.4605,…
Patients implanted with a pocket adaptor (Model 64001 and/or 64002) are limited to HEAD ONLY MRI eligibility.
FDA · DeviceMedtronic NeuromodulationWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →US
Radiometer Medical ApS recalls ABL800 FLEX Model Numbers 393-800 and 393-801 The ABL800 FLEX analyzers are intended for: " In Vitro testing of samples of whole blood for…
Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting…
FDA · DeviceRadiometer Medical ApSWorldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →US
Embecta Corp. recalls BD Nano 2nd Gen Pen Needles 32G X 4mm (5-count sample contained in Patient Counseling Sample Toolbox -Catalog 6484755/Lot Number:…
BD Nano 2nd Gen Pen Needle was shipped close to or after the labeled expiration date of February 29, 2024. Using a device past its expiry sterility and integrity of the device cannot be guaranteed.
US
Medtronic Perfusion Systems recalls TourniKwik Tourniquet Set (CFN 79012)
Incorrect component placed in four manufactured lots of the TourniKwik" Tourniquet Set.
FDA · DeviceMedtronic Perfusion SystemsWorldwide - US Nationwide distribution in the states of MN, MA, UT, CA, and…Official notice →US
Avanos Medical, Inc. recalls MIC-KEY* Gastric-Jejunal Feeding Tube Kit, Endoscopic / Radiologic Placement, Low-Profile GJ-Tube, REF 0270xx (Non-ENFIT); and MIC-KEY*…
The IFU was missing some risk information requested by another country to be contained in the IFU pertaining to potential complications.
FDA · DeviceAvanos Medical, Inc.Worldwide distribution - US Nationwide - There was also government and…Official notice →