Palodex Group Oy recalls Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray…
It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within…
Read the official notice →- Date
- 2024-08-21
- Country
- United States
- Brand
- Palodex Group Oy
- Product
- Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray…
- Classification
- Class II
- Units
- US Nationwide-Worldwide
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
Other drugs & medical recalls in the United States
- 2024-08-21
US
Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below;…
FDA · DeviceBeckman Coulter, Inc.Worldwide distribution - US Nationwide distribution in the states of CA, FL,…Official notice →US
Defibtech, LLC recalls RMU-2000 Automated Chest Compression Device
Problem in the device's motor may cause the device to stop compressions. This may lead to a delay in therapy if the device fails to operate, and could lead to patient injury or death.
FDA · DeviceDefibtech, LLCDomestic distribution to CT, FL, NJ, SC, TN. Foreign distribution to JapanOfficial notice →US
ICU Medical Inc recalls Replacement Battery List Number SUB0000594 and SUB0000864 found in Plum A+ & Plum A+3 Infusion Systems, Pump List Numbers: 11005, 11971,…
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the…
FDA · DeviceICU Medical IncWorldwide Distribution: US (nationwide) and OUS (foreign) countries of:…Official notice →US
Taro Pharmaceuticals U.S.A., Inc. recalls Hydrocortisone 1% & Acetic Acid 2% Otic Solution, 10 mL bottle, Rx only, Mfg by: Taro Pharmaceuticals, Inc., Brampton,…
Failed Impurities/Degradation Specifications: Out-of-specification results obtained for related impurities and slightly lower than the established level of the Hydrocortisone Assay obtained during stability testing.
FDA · DrugTaro Pharmaceuticals U.S.A., Inc.Product was distributed to one distributorOfficial notice →US
FDC Limited recalls Timolol Maleate Ophthalmic Solution USP, 0.5%, Sterile, 5mL bottles, Rx only, Manufactured by: FDC Limited, Waluj, Aurangabad,…
Defective container; yellow colored spike from cap lodged in the nozzle
- 2024-08-14
US
Olympus Corporation of the Americas recalls The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannula, depending on the model/outer…
Fractures and breakages in packaging trays and Tyvek covers, which may result in a sterility breach
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of CA, DE, AU, SG, KR.Official notice →