LumiraDx recalls LumiraDx SARS-CoV-2 Antibody (Ab) Liquid Quality Control
Modification to the LumiraDx SARS-CoV-2 Antibody (Ab) test that was made outside the scope of the approved EUA.
Read the official notice →- Date
- 2022-07-27
- Country
- United States
- Brand
- LumiraDx
- Product
- LumiraDx SARS-CoV-2 Antibody (Ab) Liquid Quality Control
- Classification
- Class II
- Units
- Domestic: AR, CA, FL, MA, MD, NC, NJ, NV, NY, OK, TN, TX, & WI.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More LumiraDx recalls
- 2023-06-21
US
LumiraDx recalls SteriPack Sterile Polyester Spun Swabs (25 units/Pack)- Intended for sample collection in the intended nasal area. Catalog Number: 60566RevB
Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of…
- 2022-11-02
US
LumiraDx recalls LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Control Kit Catalog #: L016080109002 (EUA) The LumiraDx SARS-CoV-2 Quality Controls are…
Assigned vial lot numbers for the incorrect type of product. SARS-CoV-2 & Flu A/B lot numbers were incorrectly assigned to the positive and negative vial lot numbers for SARS-CoV-2 Ag LQCs within batch 2002204017,…
Other drugs & medical recalls in the United States
- 2022-08-03
US
US
- 2022-07-27
US
Stryker Instruments Div. of Stryker Corporation recalls SurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting,…
When scanning sponges out after a surgical procedure, an error may allow for a duplicate sponge to be counted without the prompt to make a case note and allowing the case to close. The software update will prevent the…
FDA · DeviceStryker Instruments Div. of Stryker CorporationThe products were distributed to the following US states: KS and ME.Official notice →US
Aesculap Implant Systems LLC recalls Aesculap Sterile disposable fleece drape, Product code: GA414, individually packaged, 50 pieces per shipping unit, product code…
Packaging seal integrity not validated resulting in a lack of sterility assurance.
FDA · DeviceAesculap Implant Systems LLCUnited States: CA, CO, GA, IL, IN, KY, MI, NC, NE, NY, OH, RI, VA and WIOfficial notice →US
Aesculap Implant Systems LLC recalls AESCULAP S4 Cervical Cross Connector L24mm, S4C CROSS CONNECTOR FIXED 24MM, product code SWT113T. Used in spinal fusion surgery.
Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement of the connectors on the rods and/or microparticle abrasion.
FDA · DeviceAesculap Implant Systems LLCUnited States Nationwide distribution in the states of CA, CO, IL, IN, MI, NC,…Official notice →US
Becton Dickinson & Company recalls BD Connecta Stopcock Without Extension Tube Catalog No. 394601 (OUS)
May have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under infusion, contamination…
FDA · DeviceBecton Dickinson & CompanyWorldwide distribution - US Nationwide distribution in the states of AK, AR,…Official notice →