Drugs & medicalClass II

LivaNova USA, Inc. recalls SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).

Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline…

Read the official notice →
Date
2025-02-12
Country
United States
Brand
LivaNova
Product
SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).
Classification
Class II
Units
US: AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More LivaNova recalls

  1. 2025-02-05
  2. US
    Drugs & medicalClass II

    LivaNova USA, Inc. recalls Brand Name: SenTiva" Product Name: VNS Therapy" SenTiva", Model 1000 Model/Catalog Number: 1000 Software Version: NA Product Description:…

    Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.

    FDA · DeviceLivaNova USA, Inc.Worldwide - US Nationwide distribution including in the states of AK, AZ, CA,…Official notice →
  3. 2022-02-09
  4. US
    Drugs & medicalClass II

    LivaNova USA, Inc. recalls Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode inner diameter) and…

    A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience…

    FDA · DeviceLivaNova USA, Inc.US: NE, MO, AR, WV, KS, MA, INOfficial notice →
All 3 LivaNova recalls →

Other drugs & medical recalls in the United States

  1. 2025-02-12
  2. US
    Drugs & medicalClass II

    Hologic, Inc. recalls The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 71738-001). The…

    Three lots of Rapid fFN Specimen Collection kit (901164, 901165, and 904850) contain Rapid fFN Specimen Collection Tubes which may be empty, underfilled, or overfilled with extraction buffer. The transport tubes are…

    FDA · DeviceHologic, Inc.Worldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →
  3. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 3 M12 System Code: (1) 722063 (2) 722221

    Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get…

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Canada,…Official notice →
  4. US
    Drugs & medicalClass II

    Oxoid Limited recalls Lev5, REF CT1587B, 50 Susceptibility discs Levofloxacin 5 ug, IVD, CE 2797

    Mislabeling; Levofloxacin Antimicrobial susceptibility disc packaging may contain Norfloxacin Antimicrobial Susceptibility discs.

    FDA · DeviceOxoid LimitedUS: VA, CA, NE, MA, FL, CA, MA, NH, WA OUS: Austria, Australia, Brazil,…Official notice →
  5. US
    Drugs & medicalClass I

    ROi CPS LLC recalls regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor…

    There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends…

    FDA · DeviceROi CPS LLCDistribution was made to MO. There was no government/military/foreign…Official notice →
  6. US
    Drugs & medicalClass II

    GE Medical Systems, LLC recalls Optima XR646 HD. X-Ray imaging system.

    The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still…

    FDA · DeviceGE Medical Systems, LLCWorldwide Distribution: US (nationwide) including Puerto Rico and OUS…Official notice →
  7. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the…

    Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2.…

    FDA · DeviceFresenius Kabi USA, LLCUS distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI.Official notice →
See all drugs & medical recalls →