Kaneka America Corp recalls LIPOSORBER System, Model: LA-15, consisting of: Apheresis machine, product# 101445 model: MA-03, and disposables (product#, model)…
Labeling change to due to the risk of severe anaphylactoid reactions including shock in patients treated with the Liposorber LA-15 LDL Adsorption Column under concomitant ACE inhibitor medication treatment.
Read the official notice →- Date
- 2022-04-20
- Country
- United States
- Brand
- Kaneka America Corp
- Product
- LIPOSORBER System, Model: LA-15, consisting of: Apheresis machine, product# 101445 model: MA-03, and disposables (product#, model)…
- Classification
- Class II
- Units
- US only: CA, NH, CT, MA, FL, DE, NY, NJ, PA, OH, MD, VA, NC, SC, GA, MS, TN,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
Other drugs & medical recalls in the United States
- 2022-04-20
US
Merit Medical Systems, Inc. recalls AccessPLUS Large Bore Hemostasis Valves, Catalog Code: 580758001, 580774001, IN8152/C, K04-01197, K05T-01863, K09-09102B, K09-10388A,…
A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in difficulty advancing the guidewire…
FDA · DeviceMerit Medical Systems, Inc.US: MN, PA, AZ, FL, IL, AK, NY, CA, TX, KY, WI, NC, NJ, VA, CO, IN, AR, MO,…Official notice →US
Cardinal Health 200, LLC recalls Cardinal Health PROTEXIS PI and PI Classic, Surgical Gloves
Product was distributed without being sterilized.
FDA · DeviceCardinal Health 200, LLCProduct distributed throughout United States and GuamOfficial notice →US
Volcano Corp recalls IntraSight Mobile IVUS Systems, Model: 797415, Part numbers: 300002018101, 300002018102
Multi-Modality Touch Screen Modules (MM-TSM), which are bedside/secondary control units, used as part of Mobile IVUS systems, are missing the following labeling information: Unique Device Identifier, Legal Manufacturer…
FDA · DeviceVolcano CorpUS: CO, GA, MA, TX, AZ, NY, PA, CA, FL, MO, OR, MD, NC, TN OUS: JP, DE, GB,…Official notice →US
Conformis, Inc. recalls iTotal Identity Posterior Stabilised (PS) Knee Replacement System Model Number: TPS-301-1113-020101
Incorrect tibial base plate implant packaged in kit
US
B Braun Medical Inc recalls Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number…
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration…
US
Preferred Pharmaceuticals, Inc. recalls Econazole Nitrate Cream, 1%, Generic for Spectazole, 30 gm tube, Rx Only, Preferred Pharmaceuticals, Inc., The Physican's…
CGMP Deviations: all products within expiry are being recalled because the manufacturing firm, Teligent Pharma, Inc.is discontinuing its stability study program.
FDA · DrugPreferred Pharmaceuticals, Inc.Product distributed in California and Florida.Official notice →