Integra LifeSciences Corp. recalls TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi"…
Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized…
Read the official notice →- Date
- 2024-08-14
- Country
- United States
- Brand
- Integra LifeSciences
- Product
- TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi"…
- Classification
- Class II
- Units
- The devices are only sold in the US. The following states have received the…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Integra LifeSciences recalls
- 2024-11-06
US
Integra LifeSciences Corp. recalls Duraform Dural Graft Implant, Size 3x3 , Model Number 80-1477US Lot CT007694
Out of specification endotoxin levels.
US
Integra LifeSciences Corp. recalls SURG PAT XRAY 1/2X2 Model/Catalog Number: 801406. The surgical patties and surgical strips are indicated for the use in protection of…
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
FDA · DeviceIntegra LifeSciences Corp.Worldwide Distribution. US Nationwide, Canada Germany United Kingdom of Great…Official notice →- 2024-10-16
US
Integra LifeSciences Corp. recalls MEDIHONEY Dressing with Active Leptospermum Honey Wound and Burn Dressing W/APPLICATOR, 1.5oz tube (Model Number 31515) and 3.5oz tube…
There is a potential for pinholes in the applicator pouch film which creates a sterility concern.
FDA · DeviceIntegra LifeSciences Corp.Worldwide - US Nationwide distribution in the states of MN, IL, UT, FL, NC,…Official notice →- 2024-07-17
US
Integra LifeSciences Corp. recalls AURORA Surgiscope System. Sterile, single use device that contains a Sheath, Obturator, and Imager.
Potential sheath damage that could lead to tissue damage and/or track hemorrhage.
- 2024-07-03
US
Integra LifeSciences Corp. recalls CUSA Excel C2600 23KHz Straight Handpiece-Indicated for use in surgical procedures where fragmentation, emulsification and aspiration…
Potential for the CUSA Excel 23KHz Straight Handpiece housing to crack, may result in delay of treatment
FDA · DeviceIntegra LifeSciences Corp.Nationwide Foreign: Argentina Australia Bangladesh Bangladesh Belgium Bolivia…Official notice →- 2024-06-19
US
Integra LifeSciences Corp. recalls CODMAN BACTISEAL EVD Catheter Set - Indicated for gaining access to the ventricles of the brain and can be used with dimensionally…
Defect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal that may compromise the product sterility.
FDA · DeviceIntegra LifeSciences Corp.Worldwide - US Nationwide distribution in the state of TX and the countries of…Official notice →
Other drugs & medical recalls in the United States
- 2024-08-14
US
W.L. Gore & Associates, Inc. recalls GORE SEAMGUARD Bioabsorbable Staple Line Reinforcement Configured for Circular Staplers, 25mm; Catalog number: 1BSGC25;
Devices were labeled with an expiration date of four years and ten days rather than the validated two years.
FDA · DeviceW.L. Gore & Associates, Inc.US Nationwide distribution in the states of CA, GA, IL, NC, OH, OK, WA.Official notice →US
Baxter Healthcare Corporation recalls Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop…
FDA · DeviceBaxter Healthcare CorporationWorldwide distribution - US Nationwide and the country of Canada.Official notice →US
Philips North America Llc recalls IntelliVue Patient Monitor MX500 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866064. Indicated for use by health…
Philips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the…
FDA · DevicePhilips North America LlcNationwide Foreign: Australia Austria Bangladesh Brazil Canada Chile China…Official notice →US
Olympus Corporation of the Americas recalls Soltive Premium SuperPulsed Laser System, Model TFL-PLS
Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →US
CareFusion 303, Inc. recalls BD PYXIS MDBNK TWR MN CR-4HH-5FM-P RxNow, REF: 169-131 containing software version 3.9.1.9
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
FDA · DeviceCareFusion 303, Inc.US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL,…Official notice →US
GE Medical Systems, LLC recalls SIGNA Explorer, with affected software versions: SV25.5, SV25.6, SV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D…