Inpeco S.A. recalls FlexLab (FLX) System. Potassium Test System. in vitro diagnostic
The FlexLab System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the…
Read the official notice →- Date
- 2025-10-15
- Country
- United States
- Brand
- Inpeco
- Product
- FlexLab (FLX) System. Potassium Test System. in vitro diagnostic
- Classification
- Class II
- Units
- US distribution to CA & NY.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Inpeco recalls
- 2025-12-03
US
Inpeco S.A. recalls FlexLab (FLX); Version: FLX-217-40;
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer…
FDA · DeviceInpeco S.A.US Domestic: MI. OUS International: Australia, Finland, France, Germany,…Official notice →- 2025-10-15
US
Inpeco S.A. recalls FlexLab X (FXX) System. Potassium Test System. in vitro diagnostic
The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the…
- 2023-12-27
US
Inpeco S.A. recalls Roche Cobas 8000 and Cobas PRO Interface Module (CO8 IM)- IVD FlexLab Automation System -a modular system designed to automate…
Firmware of the Interface Module with Roche Cobas 8000 and Cobas PRO (CO8 IM) may lead to a delayed sample tube processing.
FDA · DeviceInpeco S.A.Worldwide - US Nationwide distribution in the state of MA and the countries of…Official notice →- 2023-05-10
US
Inpeco S.A. recalls Storage and Retrieval Module (SRM)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP),…
Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either…
FDA · DeviceInpeco S.A.Worldwide distribution - US Nationwide and the countries of ARGENTINA,…Official notice →- 2021-10-20
US
Inpeco S.A. recalls FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2). (Different brand names of same product.)
The Firmware (FW) of the Automation System Interface Module to ADVIA Centaur XP/XPT may lead to an incorrect association of test results to sample ID. Improper management of the sample may bring an incorrect patient…
FDA · DeviceInpeco S.A.Domestic distribution nationwide . Foreign distribution worldwide.Official notice →
Other drugs & medical recalls in the United States
- 2025-10-15
US
Applied Medical Technology Inc recalls NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 8 F x 55, REF E NRT-06055-I The NutraGlide Nasal Feeding Tube is a…
The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at lower than expected forces.
US
B BRAUN MEDICAL INC recalls PERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Winged and 19 Ga. Closed…
Potential for the lid of the catheter connector to be in the incorrect position.
FDA · DeviceB BRAUN MEDICAL INCNationwide distribution to AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS,…Official notice →US
Inpeco S.A. recalls FlexLab X (FXX) System. Potassium Test System. in vitro diagnostic
The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the…
US
LeMaitre Vascular, Inc. recalls Artegraft Collagen Vascular Graft. Model/Catalog/Part Numbers: AG630M, AG636M, AG730M, AG740M, AG845M. The Artegraft is intended to serve…
Devices were distributed with incorrect label without CE and UKCA mark, missing patient leaflet and patient implant card.
FDA · DeviceLeMaitre Vascular, Inc.No US distribution. International distribution to Great Britian and Switzerland.Official notice →US
Beckman Coulter Inc. recalls SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only.
Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which causes a delay in obtaining patient results.
FDA · DeviceBeckman Coulter Inc.US: AL KY LA MA MD MS NC NE NJ NY OK PA TX WA WV OUS: Canada Mexico SingaporeOfficial notice →US
Abiomed, Inc. recalls Automated Impella Controller (AIC), used for left heart support blood pump, labeled as the following with corresponding Product Codes: 1.…
Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
FDA · DeviceAbiomed, Inc.Nationwide distribution. International distribution to United Arab Emirates,…Official notice →