Gentuity, LLC recalls Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that…
Read the official notice →- Date
- 2026-04-22
- Country
- United States
- Brand
- Gentuity, LLC
- Product
- Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
- Classification
- Class II
- Units
- Domestic - US Nationwide distribution in the states of CA, FL, GA, IL, MA, MN,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
Other drugs & medical recalls in the United States
- 2026-04-29
US
Hetero Labs Limited (Unit V) recalls Pantoprazole Sodium, Delayed-Release Tablets, USP, 40 mg, 1000-count bottle, Rx only, Manufactured for: Camber Pharmaceuticals,…
Discoloration: Firm received five (5) complaints stating that, "Tablets discolored darker than normal and have lighter-colored spots."
- 2026-04-22
US
PHILIPS MEDICAL SYSTEMS recalls Philips Spectral CT on Rails. Model Number: 728334.
Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of…
FDA · DevicePHILIPS MEDICAL SYSTEMSWorldwide - US Nationwide distribution in the states of Minnesota and the…Official notice →US
Medline Industries, LP recalls CENTURION Circumcision Kit, Medline Kit Number/SKU CIT7050F
On lot of Sterile Circumcision Kits was distributed without the required sterilization process and is therefore not sterile.
FDA · DeviceMedline Industries, LPUS Nationwide distribution in the state of South Carolina.Official notice →US
THERAKOS DEVELOPMENT LIMITED recalls Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not…
Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment…
US
Clariance-SAS recalls Brand Name: Elegance¿ Anterior Cervical Plate Product Name: ELEGANCE 1 LEVEL CERVICAL PLATE, ELEGANCE 2 LEVEL CERVICAL PLATE, ELEGANCE 3…
Due to a quality issue regarding the presence of burr on the finished cervical plate product.
FDA · DeviceClariance-SASU.S. Nationwide distribution in the states of CA, IL, NY and TX.Official notice →US
Integra LifeSciences Corp. (NeuroSciences) recalls Cytal Burn Matrix 10x15 cm. Product ID: BMM1015.
Increased rate of out-of-specification endotoxin results.
FDA · DeviceIntegra LifeSciences Corp. (NeuroSciences)US Nationwide distribution in the states of AL, AZ, CA, DC, FL, GA, IA, IL,…Official notice →