Drugs & medicalClass II

Cook Inc. recalls Flexor Check-Flo Introducer Ansel Modification Model Number: KCFW-7.0-18/38-45-RB-ANL0-HC Order Number/GPN: G44155 Intended to introduce…

Flexor Check-Flo Introducer Ansel Modification mislabeled. Products packaged with a label stating 6FR may contain a 7FR device and vice versa

Read the official notice →
Date
2021-12-22
Country
United States
Brand
Cook
Product
Flexor Check-Flo Introducer Ansel Modification Model Number: KCFW-7.0-18/38-45-RB-ANL0-HC Order Number/GPN: G44155 Intended to introduce…
Classification
Class II
Units
AL, AZ, CA, FL, GA, KS, MI, MO, MN, NC, NY, OK, TX Foreign: BE, CH, DE, ES,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Cook recalls

  1. 2022-09-21
  2. US
    Drugs & medicalClass II

    Cook Incorporated recalls Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWAS-35-150 ORDER NUMBER (GPN): G56176. For vascular catheter…

    Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

    FDA · DeviceCook IncorporatedNationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU…Official notice →
  3. 2022-07-20
  4. US
    Drugs & medicalClass II

    Cook Incorporated recalls Pressure Monitoring Set, Single-lumen uncoated central venous catheter. RPN C-PMS-250, GPN G02838

    The products may contain excess coating material on the outside and/or interior of the needle component. Potential adverse events that may occur if an affected product is used include a delay in the procedure, a…

    FDA · DeviceCook IncorporatedUS Nationwide. Brazil, Canada,Official notice →
  5. 2022-06-15
  6. US
    Drugs & medicalClass II

    Cook Incorporated recalls Nester Embolization Microcoil RPN GPN MWCE-18-5-2-NESTER G52733 MWCE-18-5-3-NESTER G52734 MWCE-18-7-3-NESTER-01 G47338…

    The loading cartridges included with these devices may contain a small, unintended, stainless-steel cannula.

    FDA · DeviceCook IncorporatedDomestic distribution nationwide. Foreign distribution worldwide.Official notice →
  7. 2021-12-29
  8. US
    Drugs & medicalClass II

    Cook Inc. recalls Skinny Needle with Chiba Tip. RPN 090010, 090010-ET; GPN G14254, G15290. Rigid needle available with EchoTip to provide enhanced…

    Specific lots may contain black debris on the exterior of the inner stylet and/or on the interior of the outer cannula. Local inflammatory reactions may occur if an affected product is used.

    FDA · DeviceCook Inc.Worldwide distribution - US Nationwide distribution in the states of AL, CA,…Official notice →
  9. 2021-11-24
  10. US
    Drugs & medicalClass II

    Cook Inc. recalls Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer catheter, curved-tip metal…

    Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time…

    FDA · DeviceCook Inc.Domestic distribution nationwide. Foreign distribution worldwide.Official notice →
  11. 2021-11-10
  12. US
    Drugs & medicalClass II

    Cook Inc. recalls Flexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 The introducer sheath…

    Affected lots may be manufactured incorrectly with the radiopaque marker band located just below the Check-Flo proximal fitting instead of the distal tip. The issue may not be identified until the device is placed in…

    FDA · DeviceCook Inc.Domestic distribution to AR AZ CT FL GA IL LA MA ME MN ND NJ NY OH PA TX WA.…Official notice →
All 20 Cook recalls →

Other drugs & medical recalls in the United States

  1. 2021-12-29
  2. US
    Drugs & medicalClass II

    Medline Industries Inc recalls Hand Sanitizer Foam (ethyl alcohol), 62% v/v, packaged in a) 1.5 Oz. cans, NDC 53329-970-12; b) 8 Oz. (227 g) cans, NDC 53329-970-08; and…

    Defective Container: Customer complaints for leaking bottles and dispensing issues.

    FDA · DrugMedline Industries IncNationwide in the USA, Canada, Hong Kong, and PeruOfficial notice →
  3. 2021-12-22
  4. US
    Drugs & medicalClass II

    Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Medical Device Identification Cards associated with Sprint Quattro Lead, Product/CFN 6935M72

    The back of some Medical Device Identification Cards provided to patients were printed with an erroneous statement stating the following: "This patient has a complete MR conditional system implanted."

    FDA · DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)US, Canada, Northern Mariana IslandsOfficial notice →
  5. US
    Drugs & medicalClass II

    Xstrahl Limited recalls Treatment applicator for Gulmay Medical 150 or Xstrahl 150 superficial X-ray therapy system

    Base plate may detach from the main body of the treatment applicator.

    FDA · DeviceXstrahl LimitedUS and ALGERIA, AUSTRALIA, AUSTRIA, BULGARIA, CANADA, COSTA RICA, CROATIA,…Official notice →
  6. US
    Drugs & medicalClass II

    Activbody, Inc. recalls Box Label: activ5 TINY GYM IN YOUR POCKET activbody Model: Activ5 User Manual: ACTIV5 USER MANUAL Welcome to Activ5, the first…

    Due to: 1) Manufacturing issue that result in inaccurate force measurement which may lead to may potential lead to incorrect treatment planning. 2) Firm failing to register and list its medical devices with U.S. Food…

    FDA · DeviceActivbody, Inc.U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, LA MA,…Official notice →
  7. US
    Drugs & medicalClass II

    CryoLife, Inc. recalls BioGlue Surgical Adhesive, Model BG3510-5-G

    Product is not approved for use in the EU and should not have been distributed in the EU.

    FDA · DeviceCryoLife, Inc.Distribution in the European Union only (Germany/Switzerland)Official notice →
  8. US
    Drugs & medicalClass II

    Clark Laboratories, Inc. recalls Bartels Legionella Urinary Antigen- an adjunct to culture for the presumptive diagnosis of past or current Legionnaire's Disease by…

    Contamination of the conjugate may cause a pipetting issue which then could lead to an invalid assay

    FDA · DeviceClark Laboratories, Inc.OH, UTOfficial notice →