Drugs & medicalClass II

BALT USA, LLC recalls Neurovascular embolization device. The Optima Coil System is a series specialized coils that are inserted into the vasculature under…

Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the…

Read the official notice →
Date
2024-02-14
Country
United States
Brand
BALT
Product
Neurovascular embolization device. The Optima Coil System is a series specialized coils that are inserted into the vasculature under…
Classification
Class II
Units
U.S.: AZ, CA, CO, FL, GA, HI, IL, KS, MA, ME, MO, MS, NC, NE, NH, NJ, NY, OH,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More BALT recalls

  1. 2025-11-05
  2. US
    Drugs & medicalClass I

    BALT USA, LLC recalls Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast Model/Catalog Number: MEGABT09180, MEGABT09190, MEGABT091100,…

    Due to an identified manufacturing defect involving coating material lifting within the catheter's distal inner diameter/lumen.

    FDA · DeviceBALT USA, LLCUS distribution to states of: AL, CA, CO, CT, FL, IL, KY, MA, NY, TN, TX, and…Official notice →
  3. 2025-06-25
  4. US
    Drugs & medicalClass II

    BALT USA, LLC recalls Prestige Coil System REFs/UDI-DI codes: PRES0153CXPPLT 00810068567311 PRES1045CPKPLT 00810068567663 PRES1245CPKPLT 00810068567694…

    Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.

    FDA · DeviceBALT USA, LLCWorldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →
  5. 2022-10-26
  6. US
    Drugs & medicalClass II

    BALT USA, LLC recalls Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.

    Product pouch label does not match up with carton label

    FDA · DeviceBALT USA, LLCWorldwide distribution - U.S. Nationwide distribution in the states of FL, GA,…Official notice →
  7. 2022-09-14
  8. US
    Drugs & medicalClass II

    BALT USA, LLC recalls Optima Coil System Model OPTI0520CSF10. For endovascular embolization.

    Products were released to customers prior to the approval and release of a test report authorizing the use of an existing material from an alternate supplier.

    FDA · DeviceBALT USA, LLCU.S.: AZ and FL O.U.S.: NoneOfficial notice →
  9. 2022-03-02
  10. US
    Drugs & medicalClass II

    BALT USA, LLC recalls The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal…

    Due to Label/Pouch mix-up. Pouches incorrectly labeled. Incorrect shelf carton label or pouch label due to inadequate manufacturing line clearance. As a result of this labeling mix-up, there is the possibility of…

    FDA · DeviceBALT USA, LLCWorldwide distribution - U.S. Nationwide distribution in the states of FL and…Official notice →
All 6 BALT recalls →

Other drugs & medical recalls in the United States

  1. 2024-02-21
  2. US
    Drugs & medicalClass III

    Teva Pharmaceuticals USA, Inc recalls Nortrel (norethindrone and ethinyl estradiol tablets USP) 0.5/35, packaged in cartons, each carton contains 3 blister cards, each…

    Discoloration: discolored tablets (shades of blue) mixed in with the white inert remainder tablets.

    FDA · DrugTeva Pharmaceuticals USA, IncnationwideOfficial notice →
  3. 2024-02-14
  4. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS recalls Spectral CT 7500: Software Version 5.0, Model 728333 and Spectral CT 7500 China: Software Version 5.0, Model 728340.

    A software issue that can incorrectly label all captured images that may lead to misdiagnosis and subsequent incorrect treatment.

    FDA · DevicePHILIPS MEDICAL SYSTEMSWorldwide - US Nationwide distribution in the states of AR, AZ, GA, HI, IN,…Official notice →
  5. US
    Drugs & medicalClass II

    Maquet Medical Systems USA recalls Bubble Sensor (REF: 70105.5720)

    The Bubble Sensor optional accessory to the CARDIOHELP-i extracorporeal life support system includes an incorrect UDI Device Identifier (DI) provided with the label attached to the device.

    FDA · DeviceMaquet Medical Systems USAAZ, CA, CO, FL, GA, IA, IL, MN, NE, PA, TX, WA and WI.Official notice →
  6. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD BBL Sensi Disc Piperacillin/Tazobactam- 100/10 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231691…

    There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic…

    FDA · DeviceBecton Dickinson & Co.Domestic distribution nationwide. International distribution worldwide.Official notice →
  7. US
    Drugs & medicalClass II

    Angiodynamics, Inc. recalls PulseSpray Infusion System devices including: PULSE SPRAY 5F/135/20, UPN: H787125004045, REF: 12500404; Pulse Spray 3Fx135x20, UPN:…

    Rotating adaptor O-ring manufacturing defect impacting seal integrity which may lead to a leak during infusion of medication.

    FDA · DeviceAngiodynamics, Inc.United States (MD, LA & UT), Austria, Canada and Spain.Official notice →
  8. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Water Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope reprocessor for prevent…

    Water filter may have been assembled incorrectly, resulting in unfiltered tap water being used for reprocessing. This may expose a patient to a potentially contaminated device, presenting a risk of infection or sepsis.

    FDA · DeviceOlympus Corporation of the AmericasDomestic US distribution nationwide.Official notice →
See all drugs & medical recalls →